Data Input Limits
14-2
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Parameter Limits
The COBE Spectra Apheresis System operating parameters and some of
the calculated variables have maximum and minimum limits. For each
parameter, you may set up to six different maximum and minimum
limits. The limits are as follows:
•
Warning limits
•
Safety limits (for Automatic mode)
•
Absolute limits (for Manual mode)
When any parameter reaches its warning limit (in Automatic or Manual
mode), the Spectra system beeps and displays a warning message,
indicating the system is operating outside of the recommended limits.
When the Spectra system is in Automatic mode, you cannot change a
parameter beyond the safety limit. If you enter Manual mode, you may,
if absolutely necessary, enter a parameter value outside the safety limit.
Under no circumstances, even in Manual mode, will the Spectra
system exceed a parameter’s absolute limits.
The safety and absolute limits apply if you change the parameter or if the
Spectra system adjusts the parameter in response to a change you make.
For example, if you change the inlet pump flow rate, and the resulting
donor/patient platelet post-count would be below its safety limit, the
system displays a warning.
As another example, if you change the inlet pump flow rate and the
resulting AC pump flow rate would exceed the safety limit for AC
infusion, the Spectra system automatically reduces the inlet pump flow
rate to maintain the AC pump flow rate at a safe AC infusion limit, and
displays a warning.
The parameters fall into four categories:
•
Parameters you enter during donor/patient data entry (see
Table 14-1)
•
Parameters you can change during Run mode
•
Values you can configure before beginning a procedure
•
Software parameters the Spectra system calculates
Table 14-1:
Donor/patient data entry limits
Parameter
Initial Setting (Default)
Minimum Limit
Maximum Limit
Height, ft, in, or cm
(no default)
12 in or 30 cm
84 in or 220 cm
Weight, lb or kg
(no default)
10 lb or 5 kg
500 lb or 230 kg
Donor/patient initial Hct,%
*
45% for male
41% for female
10%
70% (all but RBCX)
80% (RBCX)
Donor platelet pre-count, µL
250,000/µL
1,000/µL
2,000,000/µL
Target collection bag volume, mL
(no default)
10 mL
9999 mL
Target platelet concentration in collection bag, µL
1,400,000/µL
100,000/µL
8,000,000/µL
Target run time, min
100 min
10 min
999 min
Inlet pump flow rate, mL/min
Inlet:AC Ratio x QAC
10 mL/min
+
150 mL/min
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...