Secondary Control Panel
COBE 2991 Cell Processor
•
Essentials Guide
3-9
Operator Controls
3
Secondary Control Panel
The secondary control panel (Figure 3-1) is located on the lower right side of the
COBE 2991 and is accessible by opening the right panel door. The secondary
control panel has the following controls:
•
PUMP RESTORE RATE control (in mL/min)
•
CB1 through CB5 circuit breakers
•
AUDIBLE ALARM ON/OFF switch
•
RED CELL OVERRIDE control (in seconds)
Figure 3-1: Secondary Control Panel (serial numbers 2000 and above)
PUMP RESTORE RATE (in mL/min)
The PUMP RESTORE RATE control sets the rate at which the hydraulic pump is
driven in the reverse (downward) direction to its home (starting) position. The
hydraulic pump returns to its home position during the Agitate/Wash-In operation,
or after the STOP RESET button is pressed (if the pump is not already in its home
position). This can affect the minimum agitate time, because the COBE 2991 will
not advance from the Agitate/Wash-In operation until the hydraulic pump is at its
home position. On COBE 2991s, serial numbers 2000 and above, the rate is
adjustable between 0 and 700 mL/min. On COBE 2991s, serial numbers lower
than 2000, the rate is adjustable between 0 and 450 mL/min.
Warning:
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...