AK 98 Dialysis Machine - Attentions
B:85
510 Blood pump overloaded
510 Blood pump overloaded
Check blood pump for obstruction. To continue
press Confirm.
Appears:
When the blood pump is overloaded and can hardly run in
functional check.
Machine actions:
The blood pump stops.
Operator action:
1.
Check that the blood pump segment is correctly placed,
see Section 4.2.4
“Attach the arterial blood line”
on
page A:79 in Operators handbook.
2.
Press
Confirm
to continue.
If the attention appears again, contact an authorised
technician.
Mute time:
Permanent.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...