STEP 6
STEP 5
STEP 5
STEP 5
STEP 4
healthcare
STEP 4
STEP 6
How to position Violet Room Sanitiser correctly - Part 2
• Check that the machine cycle is complete and all four lamps
show OK.
• Place the device in the room as shown.
• Move the bed away from the wall and roll the mattress to a
45°angle, lean on the left bed rails.
• Place the pillow(s) on the right bed rail in portrait position.
• If applicable, close bathroom door.
• Turn the bedside locker to face the wall as shown and
open the door/drawers.
• Move the high touch devices (including telephone if present) to a
new position to maximise exposure.
• Set the device for 10 minutes and leave the room.
• Check that the machine cycle is complete and all four lamps show OK.
• Move the device to a new location as in the step 5 image.
• Raise the bed to its highest level and raise the cot sides.
• Lean the mattress on the right bed rail.
• Place the pillow(s) in a portrait position with exposed end down, flip so the
unexposed side is facing towards the machine, lean on the left bed rail.
• Raise the patient table to its highest level and place as shown.
• Rotate the bedside locker.
• Ensure the nurse call button is placed in a position to maximise
exposure to UV e.g. dangle over side of the bed.
• Place all high touch devices e.g. TV remote controls, in a position
to maximise exposure.
• Set the device for 10 minutes and leave room.
• Once the machine cycle is complete, and all four lamps show OK,
enter the room.
• It will smell of ionization which is normal.
• Remove the UV unit from the room and place back in the UV cover.
• Use PPE if indicated by local policy and risk assessments.