The initial installation of the analyzer system is usually completed by the
purchaser. The start-up and commissioning is usually performed by
Galvanic Applied Sciences Inc. personnel.
2.1 Receiving the System
When the system arrives, inspect the packaging for external signs of damage. If there is any
obvious physical damage, contact the shipping agent and Galvanic Applied Sciences to
report the damage and request that the carrier's agent be present when the unit is unpacked.
It is recommended that you retain the shipping container so that it may be used for future
shipment of the unit, if necessary.
2.2 Environmental Requirement
2.2.1 Electrical Requirements
The power input is 100 – 240 Vac, 1 phase, 50/60 Hz, 800 W. The operating voltage is
specified on the serial number name plate
2.2.2 Temperature
The system is designed to be operated at ambient temperatures from 10 – 50
o
C. Galvanic
offers complete analyzer shelters from sun shades to complete building, please contact
Galvanic Applied Sciences, Inc. (or your local representative) for additional information.
2.2.3 Space Requirements
The system includes two cabinets, each are 30” (76 cm) H x 24” (61 cm) W x 12” (30 cm) D.
The overall outer dimension of the system with the mounting frame is 41.75” (106 cm) H x 57”
(145 cm W) x 16” (41 cm) D. The weight of the system is approximately 275 lb (125 kg), the
actual weight will depend on the system.
2.2.4 Instrument Air
The system requires 22 SCFM (37m
3
/hr) at 80-100 psi (37m
3
/hr 5.5 – 8barg).
943-TGX-CE
Refer to the 943-TGX-CE User Manual Sections 2.1, 2.2, 2.3, 2.4, 2.5, and
2.6.
Summary of Contents for 943-TGX
Page 48: ...Revision 4 48 August 02 2019 Figure 4 13 Network Sub Panel...
Page 94: ...Revision 4 94 August 02 2019 Figure 9 3 Control Cabinet Detail...
Page 96: ...Revision 4 96 August 02 2019 Figure 9 5 AC Wiring Diagram...
Page 97: ...Revision 4 97 August 02 2019 Figure 9 6 DC Signal and Wiring Diagram...
Page 98: ...Revision 4 98 August 02 2019 This Page Intentionally Blank...