4.6 Spectrum Panel
The
Spectrum
panel shown in Figure 33 shows the current
transmission spectrum
being
received by the control computer from the analyzer’s spectrometer.
Figure 33: Spectrum Panel
The transmission spectrum shows the intensity of ultraviolet light, in A/D counts, being received
at every pixel of the spectrometer. The spectrometer has a total of 2048 pixels, with an
approximate 0.3nm resolution between the pixels. The x-axis of the spectrum is unlabelled, but
it shows the wavelength range with the shortest wavelengths at the left and the longest
wavelengths on the right. Only a range of 530 of these pixels is actually used for calculating the
concentration values. The higher the A/D counts are for a given wavelength, the greater the
intensity of the light transmitted through the measurement cell to the detector at that specific
wavelength is. In the bottom right corner of the panel are a series of three fields giving data
about spectrometer parameters, shown in Figure 34.
Figure 34: Spectrometer Parameters
These three parameters are described in Table 11.
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...