background image

Gait Up SA 

EPFL Innovation Park, Bat C. CH-1015 Lausanne 

www.gaitup.com 

TVA CHE-145.136.760 

 

 

PhysiGait user manual v1.3.1 Date: 29.04.2019 
Copyright © 2018, All Rights Reserved -  Gait Up SA. CH-1015 Switzerland 

 

 

 

 

1.  PhysiGait Live elements 

 

2.  Gait recording with Physilog

®

5 from tablet 

– Basics 

 

 

Attach

 

 

Fix the sensors on the feet/ 

shoes using the rubber clips 

or foot attachment straps. 

Start 

Select options and start 

the

 

recording from the 

application. Sensors shine 

green and start to blink. 

 

 

 

 

Connect

 

Enter login information, 

go to Gait analysis and 

select “New measure”. 

Connect the 2 Physilog

®

 5 

to the application. 

Walk

 

Wait until the two sensors 

blink green synchronously 

and ask the subject to walk 

according to your protocol. 

 

 

Stop

 

Stop the recording from  

the application, the LED 

turns orange and twinkles 

while sending remaining 

data to the tablet. 

 

 

 

 

Display report

 

Once the files are 

transferred to the 

application and from there 

to the server, the Gait report 

is available. 

Summary of Contents for PhysiGait Live

Page 1: ... parameters PhysiGait Live Thank you for purchasing PhysiGait Live the easy to use system to accurately measure spatio temporal gait performances in record time Physilog 5 sensors measure the movement in the field while a dedicated mobile application computes the relevant information to offer immediate results with in lab precision This manual introduces how to use the sensors and mobile applicati...

Page 2: ... Attachment 5 3 3 Subject information selection of options and comments 5 3 4 Start the recording 6 3 5 Walk 6 3 6 Stop the recording 7 3 7 Show and compare gait reports 7 4 Gait Up Dashboard 9 5 Practical Considerations 10 5 1 Charging Physilog 5 10 5 2 Physilog 5 configuration for gait measurements 10 5 3 Handling Maintenance 11 5 4 Troubleshoot 12 6 Gait Parameter Overview 14 6 1 Essentials 14 ...

Page 3: ...t Select options and start the recording from the application Sensors shine green and start to blink Connect Enter login information go to Gait analysis and select New measure Connect the 2 Physilog 5 to the application Walk Wait until the two sensors blink green synchronously and ask the subject to walk according to your protocol Stop Stop the recording from the application the LED turns orange a...

Page 4: ...rom the list and click on the Connect button beside the sensor name The name of the sensor contains the number which is written on the back of the Physilog 5 Troubleshoot If a sensor is not detected by the tablet shortly press the main button of the sensor to enter detection mode If the sensor can still not be connected check that it has enough battery and is not connected to another mobile device...

Page 5: ...n the shoelaces or on the outer side of the shoe on the inner side there is a risk of disturbing the gait The computed gait parameters will not be influenced by the sensor position Fix the sensor such that it doesn t move with respect to the foot shoe 3 3 Subject information selection of options and comments Upon entering the Gait analysis part of the application you can select from new or previou...

Page 6: ... The recording will start shortly after sending the start signal during a short synchronisation period the app ensures that sensors receive synchronisation signals and when the countdown 3 2 1 Go has arrived at Go you can instruct the subject to start to walk according to your protocol The displayed stopwatch will start at Go and stop when you press Stop 3 5 Walk Gait Up s PhysiGait Live system ca...

Page 7: ...of the next measurement 3 7 Show and compare gait reports After the test and analysis of the data a gait analysis report can be displayed on the tablet by clicking on View Report It is also possible to initiate another measurement using the New Measure button If the results show some irregularities or values outside of normal range a warning will be displayed on the screen If an error appeared dur...

Page 8: ...ress between tests You can print and share reports from the printer and share icon in the upper right corner of the report display view in the app To print you need to have a printer connected to your mobile You can share the report via several applications email gmail dropbox etc depending on available services on your mobile Additional application features Other functions such as visualizing the...

Page 9: ...sername and password as in the mobile application Manage your Subjects Select one subject from the list to display the information in the Details window Click on modify to have access to change the information related to the patient We recommend to not enter the real name of the tested person but rather an anonymised identification code for data protection Display and print reports Select the Repo...

Page 10: ...e Physilog Plug the USB connector to a Computer or Charger voltage specifications 5V 100mA Physilog 5 must be turned off for charging Do not charge Physilog when room temperature is outside of range 0 C to 45 C Low battery level is indicated by orange blinking during measurement or if the Physilog blinks rapidly 3 times red when starting and doesn t start the recording When the Physilog is dischar...

Page 11: ... level with the App Physilog 5 can be used outside Do not use Physilog 5 to measure water sport activities without additional waterproofing Storage Store in a cool 0 C 30 C and dry place The provided box is the perfect place for your sensor Cleaning Sensors can be cleaned gently using alcohol wipes Before you clean the device make sure it is turned off and disconnect all cables Alcohol used in exc...

Page 12: ...ng ist DC 5V Stromaufnahme beträgt 100mA Alle an den Physilog 5 angeschlossenen externe Stromkreise sollen die Sicherheitskleinspannung und mit begrenzter Leistung Regeln erfüllen die in den folgenden Normen beschrieben sind IEC60950 1 2005 A1 2010 und EN60950 1 2006 A11 2009 A1 2010 A12 2011 Informations relatives à la sécurité Physilog 5 contient une batterie au lithium La batterie ne doit être ...

Page 13: ...rement started from application Green blinking 2 flashes per blink Measuring and sending data to application Orange blinking 2 flashes per blink Measuring connection to application to send data lost Green orange blinking 1 green 1 orange flash per blink Measuring and sending data to application with low battery Error indications 2 rapid red blinks Error 2 3 rapid red blinks Error 3 4 rapid red bli...

Page 14: ...10 gait cycles or more are analysed Asymmetry is the ratio of swing which compares the time in the air for the two feet in percent A perfect symmetry outputs a value of 0 The asymmetry is calculated as 𝑆𝐼 𝑆𝑤𝑖𝑛𝑔𝑙 𝑆𝑤𝑖𝑛𝑔𝑟 0 5 𝑆𝑤𝑖𝑛𝑔𝑙 𝑆𝑤𝑖𝑛𝑔𝑟 100 6 2 Temporal Analysis Cycle time s is the absolute recording time when the heel strikes the ground If you are recording a trial with several exercises of parti...

Page 15: ...rom the heel of a foot to the heel of the same foot one cycle after Stride Velocity m s is the forward speed of one cycle See Speed in essentials Peak Angle Velocity deg s is the maximum angular velocity during the swing phase between max heel clearance and min toe clearance Measured by the gyroscopes this parameter is a raw sensor measure which is not subject to any calculation errors It can be u...

Page 16: ...luate only straight walking 6 5 Clearance Analysis Note that clearance parameters are accurate when shoe size not foot size is entered in the menu of the software prior to performing data analysis Clearance calculation is based on a biomechanical model and may not be estimated in case of severe gait pattern alterations shuffling etc See Mariani et al 2012 and Dadashi et al 2014 A B C D A Max Heel ...

Page 17: ...Gait Up shall have no liability for any consequential incidental or special damages by reason of any act or omission or arising out of or in connection with the equipment or its rental delivery installation maintenance operation performance or use including without limitation any loss of use lost revenue lost profits or a cost associated with downtime The obligations contained in this paragraph co...

Page 18: ...t appareil est conforme aux niveaux limites d exigences d exposition RF pour la population globale définies par Industrie Canada L appareil ne doit pas être installé à proximité ou être utilisé en conjonction avec une autre antenne ou un autre émetteur FCC statement This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause h...

Page 19: ...y formula adaptations with respect to new version of application v1 1 2 Rebekka Anker 17 10 2017 1 1 0 Updates related to release of new app and firmware version GaitUp app v1 5 0 firmware v1 2 2 LED error indication not repeated 3 times Possibility to enter comments Stopwatch in app Gait Up Dashboard introduction Rebekka Anker 16 02 2018 1 2 0 Updates related to new versions GaitUp app v1 6 0 P5 ...

Reviews: