MIC TRAC MT-3000
8
2010 Gagemaker, LP
RCDMMT300010-10
Materials Needed:
•
MIC TRAC MT-3000 base unit and CPU readout
•
5/32" hex wrench
•
Seiko printer (optional)
•
Lightweight gage oil
•
Depth Micrometer
•
Gage Calibration Record
1. Turn the fine adjust knob clockwise to bring the
depth micrometer face in contact with the flat
face anvil.
2. Verify that the readout on the CPU displays
0.00000.
3. Loosen the coarse adjust lock and turn the
coarse adjust knob clockwise to slide the
movable carriage to the right.
4. Adjust the depth micrometer to the first
calibration value.
Note:
If you pass the calibration value, turn the
spindle back and approach the value again,
from the same direction. This practice will
increase the accuracy of the calibration.
If using a computer with
the MT-3000, go to page 11.
Calibration