Page 11
5
U
SING THE PRODUCT
5.1 Configuring the Device
The configuration process of the device has two parts: Alarm Configuration and Batch
Configuration.
5.1.1 Alarm Configuration
Connect the LM
Pro
T051 data logger to the PC via USB Type-A port. Open the LMViewE051
Software application and configure the alarm parameters as required as shown in figure 4.
For detailed process of application refer software help file.
Figure 4 Set up Alarm
In Alarm Configuration, there are four alarm conditions available for setting. The default
alarm settings (last set alarm settings) will be seen as shown in the figure 4.
The parameters of temperature alarm configuration are as follows:
1.
Alarm No.:
Alarm -1(A1) to Alarm-4 (A4).
2.
Alarm Type
: Set the alarm type High/Low as required.
3.
Set point:
Enter the appropriate Temperature value in °C as per required alarm
condition.
4.
Event:
Set the type of alarm event as
S
ingle /
C
umulative.
a.
Single Event:
Single event alarm occurs when a reading remains beyond
temperature and/or humidity alarm setpoint continuously for more than the
set period of time.
b.
Cumulative Event:
When the total time of reading exceeding the cumulative
alarm setpoint of temperature and/or humidity is more than set delay, it is
considered as a cumulative event.