- 16 -
9
.
Backlight Control
After power on, press
“
”
key to turn the backlight on or off, and
the backlight function is suitable for dim places. The default backlight turns
off every time you turn it on.
10
.
Alarm Setting
After power on, short press
“
”
to turn on and off the alarm function.
Short press "
SET
" key to set the resistance alarm value, press " " key to
move the cursor, press " " or " " key to change the current size, and
then press "
SET
" key to save and exit. When the measured value is greater
than the alarm critical setting value and the alarm function is turned on,
the meter displays the " " symbol and issues a
“
beep-beep-beep--
”
alarm
sound.The maximum value of the ground voltage alarm setting is 100V, the
maximum value of the ground resistance alarm setting is 3000
Ω
, and the
maximum value of the soil resistivity alarm setting is 9999
Ω
m. As shown
below:
11
.
Data Lock/Storage
After the power-on or measurement is completed, short press "
MEM
" key
to lock the current display data,and automatically numbered and storage. If
the memory is full, the meter displays the "
FULL
" symbol. As shown below:
The measurement data is 1032
Ω
. Short press
“
MEM
”
to display the data showed
that it is stored as the third group.