36
PROBLEM
PROBABLE CAUSE
CORRECTIVE MEASURES
4) Unusually high body fat readings (40-
50%)
a) Too much light present during the
reading.
a) Do not lift the Light Wand from the
subject's biceps until a "2" appears on
the screen during the first measurement
or the printer starts printing after the
second measurement.
b) Use the Light Shield to block the
external light. Always avoid
measurements under strong lights (e.g.
direct sunlight) because errors may
occur.
5) Printer Stops
a) Low batteries.
b) Bad battery contact.
c) Bad AC connection.
a) Replace batteries (see page 3).
b) Remove batteries and clean tabs with
isopropyl alcohol (rubbing alcohol).
c) Unplug the AC Adapter from the wall
socket and from the unit. Clean the
adapter contact with isopropyl alcohol
(rubbing alcohol).
Summary of Contents for 5000/XL