16
Active Shunt to apply internal 150K
ohm shunt.
Choose to adjust zero, no load
output, or Span, output under load.
Choose to make an automated or
manual adjustment to zero or span
adjustments.
Incremental adjustments are
available on a manual adjustment.
Click to activate auto adjust of
selected zero or span.
7
10
8
9
11
TIPS ON SHUNT AND ZERO ADJUSTMENT
The calculation side will provide
calculated shunt results from the
provided parameters.
All adjustments made are live adjust-
ments and do not need to be saved.
Zero adjustments cannot be made while
Shunt is active.
Span adjustments can be made from an
active shunt or from an applied load on
connected sensor.
Information for the calculation can
be found on the sensor’s calibration
certificate.
Sensor should be connected for all
adjustments.