Troubleshooting
FAULT:
The furnace door is not closing properly or is not sealing.
Over time the Kaowool in the door may become compacted slightly.
This may stop the door from sealing properly when
closed due to the micro door switch that is installed.
If this happens you can simply screw the connection
bolt in more to allow the door to seal correctly.
If this is done test the door switch is still working
correctly.
If the door is open, the power should cut from the
furnace.
When the door is shut the power will return.
FAULT:
The Furnace is not heating up or one side of elements is not heating up.
Shut furnace down and check elements for breaks.
If breaks present replace affected element. (See instructions above)
If none seen remove back panel and check the element connections are tight.
If issues continue replace element as a break sometimes is not visible.