d)
Statistic
The Statistic screen is for fast comparison of the last 10 measurements. By selection
single cells values can be compared and statistic value are calculated.
e)
Setup
On the Setup screen are all parameters, the can be changed by the user. On the left is the
navigation area, on the right the parameters.
Recommended settings
Designation
Settings
A-calibration value
0.000°C or -0.408°C
B-calibration value
-0.557°C or -0.600°C
Base value
-0.520°C (Limit EU)
beat
temperature.
-2.00°C ( -3.00°C Minimum)
coolingbath temperature
-6.50 °C
Mode
Celsius
Plateau
Plateau search: 0.4 m°C / 22s
Fixed time: 50 s
Maximum: 0.2m°C
Stirring device
see below