PREPARATION FOR USE

 

W&ϯϬϬϬW

HEPA filter fitting and checking the respirator 

Ensure that the high efficiency 
particulate (HEPA) filter is installed 
correctly as shown in Figure 6 

When fitting the filter element into 
the filter housing, observe that the 
filter label faces upwards as worn. 

Then applying firm pressure, click to 
engage the filter housing with the 
two filter latches. 

With a hood use the hood anti-
recirculation gasket fitted as shown 
in Figure 7. 

1.

Anti-recirculation gasket.

If a pre-filter (PF3000-04-004 or PF3000-04-005) is to be used, ensure this is positioned 
inside the filter housing prior to fitting the HEPA filter. 

More information in section ‘High efficiency particulate air filter replacement ‘on 
page

 

30. 

Figure 6

 

Figure 7

 

Summary of Contents for PureFlo 3000P

Page 1: ...til your PureFlo 3000P has been donned 1 Remove filter casing by pressing one of the side latches 4 Pinch push airflow adaptor into place short tabs face up 7 Adjust frame to head size using the rachet 8 Clip fasteners on frame to those inside hood 9 Stretch elastic around the airflow adaptor 3 Press filter into place label face up and reattach to frame 2 Insert foam gasket into the filter case li...

Page 2: ...h the hood attached 3 Roll up the cape to reveal the drawstring Tighten slightly to form a loose seal around your neck You should be able to fit three fingers in the gap 3 Roll down the cape Ensure the respirator is assembled correctly and donned BEFORE entering the contaminated area Long hair should be worn loose and down behind your shoulders Wash your hands thoroughly before and after donning y...

Page 3: ...e is for illustrative purposes only and should be adapted to meet your Trusts Policies 1 Roll up the cape ensuring the outside is tucked in and the toggle is visable 7 Loosen the drawstring toggle bend forward and have your doffing buddy remove the hood Clean or dispose as required 3 Locate the airflow adaptor at the rear of the hood and stretch elastic over the top 2 Release the hood from the fro...

Page 4: ...ntaining 1000 ppm available chlorine Alternatively a combined detergent disinfectant may be used Do not soak hoods 3 Rinse head Top Wipe all components cleaned with a clean soft cloth dampened with clean warm water Note It is important to rinse off chlorine based products While rinsing is preferred certain other disinfectants may not require this step Follow the user instructions for the product s...

Page 5: ...upwards as worn Then applying firm pressure click to engage the filter housing with the two filter latches With a hood use the hood anti recirculation gasket fitted as shown in Figure 7 1 Anti recirculation gasket If a pre filter PF3000 04 004 or PF3000 04 005 is to be used ensure this is positioned inside the filter housing prior to fitting the HEPA filter More information in section High efficie...

Reviews: