ELECTRICAL AND AIR PIPE LINE CONNECTIONS
●
WORK CLAMPING CYLINDER CONTROLLED BY LIMIT SWITCHES
SET
> >
DISPLAY
For further details of the test condition setting procedures, refer to page 38.
●
Operation timing is as follows:
・
Set the DLY timer to delay pressure applying timing
(
①
).
・
Error 5 and 6 functions are only controllable via REMOTE connector.
RDY
DLY
CHG
DET
EXH
RDY
START
LMT1
LMT2
SOL1
①
0.5sec
EXHSETTINGTIME
※
In the Internal Operation Mode, the tester cannot run.
SOL1
27
SOL1
28
LMT1
5
LMT2
6
COM-
9・10
COM+
11・12
①
WORK CLAMPING CYLINDER
CYLINDER
CONTROL VALVE
②
③
◎
UNCLAMPING SWITCH
CLAMPING SWITCH
ON
OFF
・
PRESET THE CYLINDER CONTROL METHOD
SWITCH
STATUS
OFF
CONTROL BY LIMIT SWITCHES
ERROR5: (UN)CLAMPING NOT COMPLETED AFTER 3 MIN.
ERROR6: NO LMT1 SIGNAL AT THE STARTING TIME
∬
∬
●
C line to
①
,
and
-
line to
②
.
For a 3-wire cylinder,
connect power line to
③
.
●
Choose limit switches that fulfill the following specifications:
・
Load (power) voltage: DC12V
・
Power Consumption (for 3-wire switch): 50mA or less for the total of the load connected to COM +
・
Load current: 5mA
・
Voltage Drop: Less than 3V
・
Leakage Current: Less than 1mA
●
Choose solenoid valves for cylinder controls that fulfill the specifications indicated below:
・
Driving Voltage
: FL-2
□□□
-0
:
AC100V
50/60HZ
FL-2
□□□
-2
:
To accord with external valve power source specs.
・
Power Consumption : Less than 100VA
・
Preset the Air Leak Tester testing condition to:
●
Set to condition setting status.