Output display
The system meets the AIUM (American Institute of Ultrasound in Medicine) output
display standard for MI and TI, see
on page 418. The
following table indicates for each transducer and operating mode when either the TI or
MI is greater than or equal to a value of 1.0, thus requiring display.
Transducer Model
Index
2D, M-Mode
Color
UHF22
MI
TIC, TIB, or TIS
UHF48
MI
TIC, TIB, or TIS
UHF70
MI
TIC, TIB, or TIS
TI or MI ≥ 1.0
Even if MI is less than 1.0, the system provides a continuous real-time display of MI in
all imaging modes, in increments of 0.1.
The system meets the output display standard for TI and provides a continuous real-
time display of TI in all imaging modes, in increments of 0.1.
The TI consists of three user-selectable indices, and only one of these is displayed at
any one time. In order to display TI properly and meet the ALARA principle, the user
selects an appropriate TI based on the specific exam being performed. FUJIFILM
SonoSite provides a copy of AIUM Medical Ultrasound Safety, which contains
guidance on determining which TI is appropriate. See
on page 418 or
"Guidelines for reducing MI and TI"
Output display
381
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...