Quick Start Guide to the Vevo 770

® 

Rev. 1.7 

Page 

5

 

EKV™ Reconstruction 

1. 

Check the ECG monitor to ensure that the system is producing a 
consistent, strong ECG signal. 

2. 

Connect the RMV scanhead to the Vevo Integrated Rail System III arm 
mount. 

3. 

Acquire a B-Mode image. 

4. 

Select Setup > B-Mode Setup, or press <Mode Setup> to change 
acquisition parameters. 

5. 

Click Start EKV or press <EKV™>. The progress bar beneath the image 
area tracks the EKV acquisition. When acquisition ends, the EKV loop has 
been acquired. 

6. 

To save the acquired EKV data, press <Cine Store>. 

3D-Mode 

1. 

Connect the 3D motor stage to the Vevo Integrated Rail System III. 

2. 

Connect the RMV clamp to the 3D motor stage. 

3. 

Connect the RMV scanhead to the RMV clamp. 

4. 

Select Setup > 3D-Mode Setup or press <Mode Setup> to change the 
acquisition parameters. 

5. 

Click Init. 3D to initialize the 3D motor stage. 

6. 

Click Start 3D. The 3D motor stage begins stepping along the acquisition 
route. 

7. 

To save the acquired 3D data, press <Cine Store>. 

Power Doppler Mode 

1. 

In B-Mode acquisition, identify the target region for Power Doppler 
acquisition. 

2. 

Select Overlay > Power Box Wire-frame or press <Overlay> until the 
Power Doppler Mode sample volume is displayed. 

3. 

Position and size the sample volume to include the target region for Power 
Doppler signal acquisition on the B-Mode image. 

4. 

Switch to Power Doppler Mode from B-Mode by selecting Mode > Power 
Doppler Mode
, or by pressing <Power>. 

5. 

Start and stop data acquisition using <Scan/Freeze>. 

Note: Power Doppler Mode is also supported in 3D-Mode.  To acquire Power-
3D-Mode data, from the Power Doppler Mode image click Start 3D. The 3D 
motor stage begins stepping along the acquisition route. 

Summary of Contents for Vevo 770

Page 1: ...nded for use on any living human being Start up Procedure Connect the RMV Attach the RMV scanhead that will be used during the imaging session to one of the connectors on the front of the Vevo 770 Note Only the 700 Series RMV scanheads are to be used with the Vevo 770 Earlier versions of the RMV scanheads are not compatible with the Vevo 770 Start the Vevo 770 1 Turn on the Vevo system using the b...

Page 2: ...oftware startup process 3 Once a scanhead is initialized the Study Browser is displayed The Study Browser is used for data acquisition and for the review and analysis of study data 4 Click New or press New and fill out at least the two mandatory study information fields Operator Name and Study Name 5 Click OK to continue with the study The Vevo 770 will automatically start acquiring data in B Mode...

Page 3: ...ughout the imaging session AM Mode Anatomical M Mode 1 In B Mode acquisition identify the target region for AM Mode acquisition 2 Select Overlay M Mode Sample Volume Wire frame or press Overlay until the M Mode sample volume is displayed 3 Right click the image area to switch to the AM Mode sample volume from the M Mode sample volume The letter A should be visible in the top left corner of the B M...

Page 4: ...gle the Scout image to verify the sample volume location throughout the imaging session Tissue Doppler Mode 1 In B Mode acquisition identify the target region for PW Doppler acquisition 2 Select Overlay Tissue Doppler Sample Volume Wire frame or press Overlay until the Tissue Doppler Mode sample volume is displayed 3 Position and size the sample volume to include the target region for Tissue Doppl...

Page 5: ... Mode Setup or press Mode Setup to change the acquisition parameters 5 Click Init 3D to initialize the 3D motor stage 6 Click Start 3D The 3D motor stage begins stepping along the acquisition route 7 To save the acquired 3D data press Cine Store Power Doppler Mode 1 In B Mode acquisition identify the target region for Power Doppler acquisition 2 Select Overlay Power Box Wire frame or press Overlay...

Page 6: ...the study and click Commit Session Data to save the data To copy the study to an accessible drive click Copy To Shutting Down the Vevo 770 1 At the end of the session click Close Study or press Close 2 Save the study data if desired by clicking Commit Session Data 3 Press the computer standby switch on the left side of the system to invoke the Windows shut down process 4 Click Turn Off from the pr...

Reviews: