7.3
EOG (Ethylene Oxide Gas) Sterilization
(1) Ensure that this product is completely dry, and
then detach the connector cap from the probe
connector.
(2) Put this product in a sterile pack, and then seal
the pack tightly. Place it in an EOG sterilizer.
(3) Sterilize this product according to the operating
procedure of the EOG sterilizer.
(4)
Sterilize this product for a specified time, and
aerate it according to operating procedure of
the EOG sterilizer.
Temperature
Time
55ºC
12 hours