6.4
Rinsing of Endoscopic Ultrasonic Probe After Disinfection
(1) Fill a sterile basin with sterile water.
(2) Fully immerse the endoscopic ultrasonic probe
in sterile water.
(3) Wipe the external surface (insertion portion,
boot, probe connector, end holder and
connector cap) of the endoscopic ultrasonic
probe with sterile gauze in sterile water to
remove air bubbles completely.
(4) Rinse the endoscopic ultrasonic probe by
gently shaking it in sterile water for 30 seconds.
(5) Remove the endoscopic ultrasonic probe from
sterile water and transfer it to a clean basin.