Guidance and manufacturer’s declaration - electromagnetic immunity
The CR-IR 355RU is intended for use in the electromagnetic environment specified below.
The customer or the user of the CR-IR 355RU should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
CR-IR 355RU, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d
= 1.2
P
d
= 1.2
P
80 MHz to 800 MHz
d
= 2.3
P
800 MHz to 2.5 GHz
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the CR-IR 355RU is used exceeds the applicable RF
compliance, the CR-IR 355RU should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the CR-IR 355RU.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
Summary of Contents for CR-IR 355RU
Page 2: ...ii FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 4: ...iv FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 40: ...4 4 Utility 4 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 50: ...5 10 Troubleshooting 5 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 56: ...6 6 Care and Maintenance 6 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 62: ...A 6 Speci cations A FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 66: ...B 4 IP Handling B FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 74: ...Y 2 Image Processing Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 122: ...2 Maintenance and Inspection FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 124: ...2 Index FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 125: ...FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Page 126: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...