6 - 11
INZ-TN4ZAFE-E
6.4 Changeover of range (When 2 ranges have been selected)
This mode is used to select the ranges of measured components.
(1) During measurement, press the
MODE
key to
display the User mode.
Point the cursor to “Changeover of
Range”. Press the
ENT
key.
To close the “Changeover of range setting
process.” or cancel this mode midway,
press the
key.
A previous screen will return.
To close Zero Calibration
ESC
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
Select an item
with UP/DOWN and ENT
Range Change
ENT : Select range
ESC : Back
H
2
Range1
Range2
0–3 vol%
0–100 vol%
Range Change
Select range
with UP/DOWN and ENT
Back with ESC
H
2
Range1
Range2
0–3 vol%
0–100 vol%
(2) On the Range Change screen that
appears, press the
ENT
key to change
over the range.
(3) Select the range you want to set by using
key and press the
ENT
key.
Measurement can be made within the
range selected.
In this case, the range identification
signal is active (ON) with the Low range
(Range 1) and inactive (OFF) with the
High range (Range 2).
Note) When the remote range is set to ON,
range changeover is not allowed on the
screen.