6 - 10
INZ-TN4ZAFE-E
6.3.2 Forced stop of auto calibration
This mode is used to cancel the auto calibration forcedly.
(1) During measurement, press the
MODE
key to
display the User mode. Point the cursor
to “Setting of Auto Calibration” by press-
ing the key. Press the
ENT
key.
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
Select an item
with UP/DOWN and ENT
About Auto Cal.
Stop Auto Calibrati
Time : MON 17 : 24
Start Time
SUN 12 : 00
Cycle
07
day
Flow Time
300 sec.
ON/OFF
OFF
Select setting item
About Auto Cal.
Time : MON 17 : 24
Start Time
SUN 12 : 00
Cycle
07
day
Flow Time
300 sec.
ON/OFF
OFF
Stop auto calibration?
Stop with ENT
Not with ESC
Stop Auto Calibrati
(2) In the “Setting of Auto Calibration” item
selection screen that appears, point the
cursor to “Stop Auto Calibration” by
pressing the key. Press the
ENT
key.
(3) “Stop Auto Calibration” is inverted.
A message appears, prompting you to
verify that you want to cancel or con-
tinue auto calibration. To cancel the auto
calibration, press the
ENT
key. If you
press the
ESC
key, auto calibration is not
stopped.