6.2.6 Command HP
<id>HP<parameter><CR>
Heating power Condensation protection
Access
:
Read / write.
Description:
The heating capacity for the condensation protection of the
precipitation monitor is set by command “HP”. The indication is stated
in percent.
Parameter description:
5…8: Maximum responsivity of the precipitation monitor for detection
of the lowest precipitation intensities.
9…17: High responsivity, and at the same time reduction of the
responsivity for fog, mist and condensation.
18…22: Maximum insensibility against fog, mist and condensation
Value range:
0...100
Initial value:
5
6.2.7 Command ID
<id>ID<parameter><CR>
Identification number
Access:
Read / write.
Description:
This command is used to specify the identification number
(INTERNER BUS interpreter) resp. the slave address (MODBUS RTU
Interpreter). A response telegram will only be transmitted if the 'id' in
the command matches the one set in the Weather Station. An
exception here is the generic 'id' that causes all weather stations to
respond. Once the 'id' has been changed, the device will immediately
respond with the new 'id'.
Parameter description:
99
generic 'id' (INTERNER BUS interpreter)
0
Broadcast slave address (MODBUS RTU Interpreter)
Value range:
0 to 99 (INTERNER BUS interpreter)
1 to 247 (MODBUS RTU interpreter)
Initial value:
0 (INTERNER BUS Interpreter)
1 (MODBUS RTU Interpreter)
6.2.8 Command KY
<id>KY<parameter><CR>
Key/password
Access:
Read / write.
Description:
Through of the command “KY” the value for the key (password) is set.
For the change of parameters the required password must be used.
Parameter description:
0
no password
234
password for user level
Value range:
0 / 234
Initial value:
0