observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional  measures  may  be 
necessary, such as reorienting or relocating the   

[EQUIPMENT or SYSTEM]

b

        Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 

 

 

Recommended separation distances between   

portable and mobile RF communications equipment and the 

 

[EQUIPMENT or SYSTEM]

The   

[EQUIPMENT  or  SYSTEM]  i

s

  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF 

disturbances  are  controlled.  The  customer  or  the  user  of  the   

[EQUIPMENT  or  SYSTEM] 

can  help  prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications 
equipment  (transmitters)  and  the 

 

[EQUIPMENT  or  SYSTEM] 

as  recommended  below,  according  to  the  maximum 

output power of the communications equipment. 

Rated maximum 

output power of 

transmitter 

(W) 

Separation distance according to frequency of transmitter 

(m) 

 

150 KHz to 80 MHz 

P

d

167

.

1

=

 

80 MHz to 800 MHz 

P

d

167

.

1

=

 

800 MHz to 2.5 GHz 

P

d

333

.

2

=

 

0.01 

0.117 

0.117 

0.233 

0.1 

0.369 

0.369 

0.738 

1.167 

1.167 

2.333 

10 

3.689 

3.689 

7.379 

100 

11.667 

11.667 

23.333 

For  transmitters rated at a maximum  output power not listed above, the  recommended separation distance d in 
metres  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the  transmitter,  where  P  is  the 
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1     At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2     These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

 

 

FCC ID: 2ADNQFTB13WBT   

This device complies with Part 15 of the FCC Rules. Operation is subject to the   

Following two conditions: 

(1)  This device may not cause harmful interference, and (2) This device must accept any 

interference received , including interference that may cause undesired operation. 

 

 

 

 

 

 

 

 

Summary of Contents for FT-B11W

Page 1: ...manual appropriately for future reference This User Manual is suitable to the following models FT B11W FT B12W FT B13W FT B14W FT B21Y FT B22Y FT B11W V FT B12W V FTB13W V FT B14W V FT B21Y V FT B22Y...

Page 2: ...V FT B12W V FTB13W V FT B14W V FT B21Y V FT B22Y V FT B13W UR for UART port connection FT B13W BT for Bluetooth connection Measurement Principle Oscillography Cuff Soft cuff Cuff size 290mm 72mm 5mm 1...

Page 3: ...OFF Memory or 3 ON OFF Set Memory Life Time Machine 5 years or 10000 times Cuff 10000 times Contents 2 x 1 5V AAA alkaline batteries 3V Optional Storage case Instruction Manual Contraindications 1 Hea...

Page 4: ...n intended operator The patient can measure transmit data and charge battery under normal circumstances and maintain the device and its accessories according to the user manual 3 If the cuff pressure...

Page 5: ...16 If you use a cardiac pacemaker consult your physician before using the device 17 Be sure to use this device only for measuring blood pressure DO NOT use it for any other purpose 18 DO NOT use this...

Page 6: ...d rub when clean the cuff 7 DO keep the device out of reach of children pets and insects III Name of Each Part Figure 1 Appearance LCD Display SET Button Calendar Clock Setting Mode MEMORY Button Cale...

Page 7: ...shing down on arrow and sliding cover forward 1 Date Month Day 2 Time Hour Minute 3 Systolic Blood Pressure unit mmHg 4 Diastolic Blood Pressure unit mmHg 5 Pulse Rate unit beat minute 6 WHO Blood Pre...

Page 8: ...avoid damage from the possibility of leaking batteries All the measurements will remain in the memory should the batteries become drained removed or replaced Figure 3 Figure 4 Clock Adjusting and Uni...

Page 9: ...sing a 24 HOUR clock AM PM is not displayed NOTE WHEN EVERYTHING IS SET UP COMPLETELY THE MONITOR WILL AUTOMATICALLY SWITCH OFF BUT WHEN YOU PRESS ON OFF BUTTON AGAIN THE MONITOR WILL ACTIVATE Figure...

Page 10: ...uff should be at approximately the same height as your heart TIPS FOR BLOOD PRESSURE MONITORING Relax for about 5 minutes before measurement Do not smoke or ingest caffeine at least 30 minutes prior t...

Page 11: ...play screen See Figure9 NOTE Do not self diagnosis according to measured result Consult with your physician for further diagnosis If the device cause any discomfort during measuring process or fail to...

Page 12: ...announced Figure 8 Figure 9 Assessing High Blood Pressure for Adults The follow standards for assessing high blood pressure without regard to age or gender have been established as a guideline accordi...

Page 13: ...nsion 140 159 90 99 YELLOW Stage 2 Hypertension 160 179 100 109 ORANGE Stage 3 Hypertension 180 110 RED Note The graph is not exact but may be used as a guide in understanding non invasive blood press...

Page 14: ...models FT B11W V FT B12W V FT B13W V FT B14W V FT B21Y V and FT B22Y V UART connection and Blue Tooth Function The UART connection model is FT B13W UR UART Universal Asynchronous Receiver Transmitter...

Page 15: ...NIBP Start to measure according to the normal blood pressure monitor operation method After measuring the result will be displayed on LCD including systolic pressure diastolic pressure and pulse will...

Page 16: ...tening which means it will be status of Listening once open As shown in Figure 2 and Figure 3 Figure 2 Figure 3 Then go to Host Slave in Write new value input below commands respectively Connect blood...

Page 17: ...ality Reason Checkout LCD shows Low Battery icon Batteries are low Change new batteries Shows abnormal result Cuff is not tightened properly or its position is incorrect Tighten cuff correctly and ref...

Page 18: ...tivated then stop using it Cuff inflation rate is too low or does not inflate Cuff or bladder inside the cuff leakage air Purchase a new cuff Cuff deflates too quickly Cuff has been applied too loose...

Page 19: ...t 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequenc...

Page 20: ...f floor are covered with synthetic material the relative humidity should be at least 30 If ESD interfere with the operation of equipment counter measurements such as wrist strap grounding shall be con...

Page 21: ...800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field str...

Page 22: ...rding to frequency of transmitter m 150 KHz to 80 MHz P d 167 1 80 MHz to 800 MHz P d 167 1 800 MHz to 2 5 GHz P d 333 2 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689...

Page 23: ...PART Symbol for Follow operating instructions IP22 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Managemen...

Reviews: