8. Instruction for Use

Rev. Date

07.09..2017

Page

6/13

display the average value of the latest 14 tests. (The calculation could be made only when there are 14 or
more testing values)

3

Maintenance after using: Do NOT damp the tester and wipe off any liquid at the insert and inside; it should use

the soft cloth to wipe off the dirt on the surface; and the product should be stored under the normal
temperature, do NOT place it in the electromagnetic environment or fall it down.

4

Please take out the batteries if it’s not being used for a long time.

5

Unit: mmol/L or mg/dL

1mmol/L

18mg/dL

1. The symbol of displayed on LCD in the measurement

(a) “888”: self-checking;
(b) When “Bottle” displays and blinks, it’s required to insert the Code card;
(c) When “Test Strip” displays and blinks, it’ required to insert the test strip;
(d) When the symbol of “blood” displays at the top side, it’s required to get the blood;
(e) After getting the blood, it shall count down and display at the left side of LCD, the tester enters the self-testing

state;

(f) The result shall be displayed in XXX

mmol/L

or XXX

mg/dL

; when the result

2.2mmol/L

40mg/dL

,

LCD shall display “LO”, and then the result >27.8 mmol/L(500mg/dL), LCD displays “HI”;

(g) When LCD displays “Err”, it means the test has error;
(h) When the “

“ symbol displays, it means the batteries is used up and should be replaced;

(i) When the “thermometer” symbol displays at the upper side of bottle on LCD, it means the ambient

temperature is higher than the normal working condition, it should use the tester in the ambient temperature of
10

-40

(50 °F

104 °F).

(j) When the “month”, “day” and “time” symbol display, it means the date and time of the test.

1. Identification of error information and measures

1

“Err”—it means the test has error.

Possible reasons:

the operating method (step) is wrong; invalid test strip is used; the blood is not enough; the

blood is added repeatedly; the ambient temperature is not in the range of 10

-40

(50°F

104 °F)the test strip is moved in testing; the tester has fault(s).

Suggestion:

Please read the operating manual carefully to use the blood glucose monitor

correctly and make the test once again by the correct method.

Please

contact our after-sale service department.

2

“LO”—it means the measuring

value is lower than 2.2 mmol/L

40mg/dL

; “HI”—it means the measuring value

is higher than 27.8 mmol/L (500mg/dL).

Suggestion:

Check whether the operation is correct by doing the test once again with the
effective test strip, if the value is not changed.Please consult your physician.

Please contact our after-sale service department.

Summary of Contents for eB-D11

Page 1: ...for Use Rev Date 07 09 2017 Page 1 13 Instruction for Use Blood Glucose Monitor and Blood Glucose Test Strip eB G BT eB D12 BT eB D33 BT eB D34 BT eB D43 BT eB D44 BT eB D45 BT eB D52 BT eB G eB D11 e...

Page 2: ...s 5 VII Description and explanation of icon on LCD 6 VIII Troubleshooting and precautions 7 VIIII Knowledge on blood glucose and diabetes 8 X Statements and Declarations 9 Hand hold quick blood glucos...

Page 3: ...ingle package has been opened for three months and do NOT use them 2 Introduction to the package of test strip It should finish the test within five minutes after the test strip is opened the test str...

Page 4: ...mW 3 Type of test strip Siphon type 4 Parameters of test strip a Dimensions 32mm 8mm 0 8mm L W D b Nominal weight 0 5g piece The product must be used together with siphon blood glucose test strip prod...

Page 5: ...Step 7 Application of sample 1 Please apply the blood sample at the siphon mouth of reaction area when the LCD displays Blood Entry 1 Press Button Power shortly to turn on and 88 8 will be fully disp...

Page 6: ...and then the result 27 8 mmol L 500mg dL LCD displays HI g When LCD displays Err it means the test has error h When the symbol displays it means the batteries is used up and should be replaced i When...

Page 7: ...ll b The initial blood sample is not enough measuring the test after adding the blood again c The blood sample is polluted d The test strip is pulled out and then inserted e It s stored in the low tem...

Page 8: ...lysaccharide could not be called blood glucose until they are converted to glucose The thickness of blood glucose of a normal person should be stable and balance Reference standard of blood glucose va...

Page 9: ...r stacked with other equipment and that if adjacent or stacked use is necessary this machine should be observed to verify normal operation in the configuration in which it will be used Warning changes...

Page 10: ...G BT use RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emission CISPR 11 Class B The e...

Page 11: ...by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF...

Page 12: ...Hz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended sepa...

Page 13: ...with your local Authority or retailer for recycling advice Symbol for TYPE BF APPLIED PART Symbol for Follow operating instructions IP22 Symbol for the IP classification Symbol for RF transmitters Ma...

Reviews: