8. Instruction for Use
Rev. Date
07.09..2017
Page
6/13
display the average value of the latest 14 tests. (The calculation could be made only when there are 14 or
more testing values)
(
3
)
Maintenance after using: Do NOT damp the tester and wipe off any liquid at the insert and inside; it should use
the soft cloth to wipe off the dirt on the surface; and the product should be stored under the normal
temperature, do NOT place it in the electromagnetic environment or fall it down.
(
4
)
Please take out the batteries if it’s not being used for a long time.
(
5
)
Unit: mmol/L or mg/dL
(
1mmol/L
=
18mg/dL
)
1. The symbol of displayed on LCD in the measurement
(a) “888”: self-checking;
(b) When “Bottle” displays and blinks, it’s required to insert the Code card;
(c) When “Test Strip” displays and blinks, it’ required to insert the test strip;
(d) When the symbol of “blood” displays at the top side, it’s required to get the blood;
(e) After getting the blood, it shall count down and display at the left side of LCD, the tester enters the self-testing
state;
(f) The result shall be displayed in XXX
(
mmol/L
)
or XXX
(
mg/dL
)
; when the result
<
2.2mmol/L
(
40mg/dL
)
,
LCD shall display “LO”, and then the result >27.8 mmol/L(500mg/dL), LCD displays “HI”;
(g) When LCD displays “Err”, it means the test has error;
(h) When the “
“ symbol displays, it means the batteries is used up and should be replaced;
(i) When the “thermometer” symbol displays at the upper side of bottle on LCD, it means the ambient
temperature is higher than the normal working condition, it should use the tester in the ambient temperature of
10
℃
-40
℃
(50 °F
~
104 °F).
(j) When the “month”, “day” and “time” symbol display, it means the date and time of the test.
1. Identification of error information and measures
(
1
)
“Err”—it means the test has error.
Possible reasons:
the operating method (step) is wrong; invalid test strip is used; the blood is not enough; the
blood is added repeatedly; the ambient temperature is not in the range of 10
℃
-40
℃
(50°F
~
104 °F)the test strip is moved in testing; the tester has fault(s).
Suggestion:
①
Please read the operating manual carefully to use the blood glucose monitor
correctly and make the test once again by the correct method.
②
Please
contact our after-sale service department.
(
2
)
“LO”—it means the measuring
value is lower than 2.2 mmol/L
(
40mg/dL
)
; “HI”—it means the measuring value
is higher than 27.8 mmol/L (500mg/dL).
Suggestion:
①
Check whether the operation is correct by doing the test once again with the
effective test strip, if the value is not changed.Please consult your physician.
②
Please contact our after-sale service department.