8. Instruction for Use
Rev. Date
07.09..2017
Page
4/13
3
③
Electrode: insert this end into the corresponding place of the blood glucose tester to start the device,
please keep the blue arrowhead side upwards in inserting.
1.Testing scope: 2.2 mmol/L
~
27.8mmol/L (40mg/dL
~
500mg/dL).
2.Testing time: ≤10s.
1.Model No.: eB-G-BT
(Suitable for eB-D12-BT,eB-D33-BT,eB-D34-BT,eB-D43-BT,eB-D44-BT,eB-D45-BT,eB-D52-BT
,
eB-G,eB-D11,eB-D12,eB-D21,eB-D31,eB-D32,eB-D33,eB-D34,eB-D35,eB-D41,eB-D42,eB-D43,
eB-D51,eB-D52)
2.Parameters of blood glucose monitor:
(a) Power supply: DC 3.0V (two AAA alkaline batteries);
(b) Display mode: mmo1/L or mg/dL display and testing time display;
(c) Times of result memory: 180 times;
(d) Dimensions: 80mm*58mm*23mm ;
(e) Weight: 46g
(f) Rated power: 30mW.
3.Type of test strip: Siphon type
4.Parameters of test strip
(a) Dimensions: 32mm×8mm×0.8mm (L*W*D)
(b) Nominal weight: 0.5g/piece
·
The product must be used together with siphon blood glucose test strip produced.
·
Blood glucose monitor is applicable for out-of-body test only.
·
Blood glucose monitor is applicable for the test of ending whole blood.
Ambient conditions (normal working conditions)
(a) Ambient temperature: 10
℃
-40
℃
(50 °F
~
104 °F);
(b) Relative humidity: ≤80%;
(c) Atmospheric pressure: 700hPa
~
1060hPa;
(d) Power supply: DC 3.0V (two AAA alkaline batteries);
1. Use the blood glucose monitor to test the blood glucose. Step
1: adjustment of time
Power On: Press Power switch for five seconds to enter the setup interface of time. Firstly the Year place shall
blink, press Memory key to adjust the year and then further press Power key to enter the adjustment interface of
month, use the same operation to finish the adjustment of day, hour and minute.
Step 2: Installation of pinhead
(1)Open the protective cover of blood taking device;