Selecting a Process Fluid
This recirculator is designed for operation with a host of industry standard fluids. The
fluid you choose will depend on what temperature range is required. Ethylene Glycol
and water (60/40 mix) is an inexpensive and readily available fluid for
–25°C to +60°C.
For a fluid outside this temperature range, SP Scientific suggests using HT-30
Silicone Oil for
–60°C to +100°C.
There are other fluids available, but to avoid potential application problems and to
speed troubleshooting; it is suggested that one of the fluids mentioned above be used.
DO NOT USE AUTOMOTIVE ANTI-FREEZE (it contains clotting agents). If for some
reason one of the above fluids is not suitable, and you select a different one, be sure
that it has a viscosity of less than 20 centistokes over the full temperature range of the
unit.
Note:
All of the above mentioned fluids have viscosity of less than 20 centistokes across their
useable temperature ranges.
Since most systems have capability to achieve temperatures below 0°C, water should
not be used as a process.
Fluid
Temperature
Range (°C)
Flammability
Toxicity
Dielectric
Ethylene Glycol/Water (60/40 mix)
-25 / +60
Low
Low
Low
HT-30 Silicone Oil (Recommended)
-60 / +100
Low
Very Low
High
Summary of Contents for MAXI-COOL
Page 1: ...RECIRCULATING CHILLER OPERATOR S MANUAL MAXI COOL with EDC Control ...
Page 2: ......
Page 6: ...iv Rev 007 04 12 SP Scientific 2012 ...
Page 10: ...Introduction Maxi Cool 2 Rev 007 04 12 SP Scientific 2012 ...
Page 19: ...Maxi Cool Installation and Startup Rev 007 04 12 11 SP Scientific 2012 ...
Page 26: ...Operation Maxi Cool 18 Rev 007 04 12 SP Scientific 2012 ...
Page 36: ...User Menu Maxi Cool 28 Rev 007 04 12 SP Scientific 2012 ...
Page 65: ......