FSI Calibration Guide
Using CR100 / ColourSpace / BoxIO
© 2022 Flanders Scientific, Inc.
3
Now, click on the Probe Options button in
ColourSpace. From the Probes menu select
Colorimetry Research CR100. Select Rescan
if no active com ports are shown. Then press
connect. Once connected verify you have the
correct monitor matrix active on the probe. We
suggest setting extra delay to at least 0.5 sec-
onds as a setting of 0 seconds will cause read
errors.
Next, ensuring that grey is still being displayed
on your monitor and the probe is positioned in
front of the screen, click on Probe Calibration to
establish sync for your CR100 and press OK.
Plug the CR100 into the
computer and aim it towards
the middle of the monitor.
Click on the Manual Measure button
and move the sliders to ensure test
patterns are being properly generated.
Set the sliders to 128,128,128 to display
grey on screen.