FSI Calibration Guide 

for AM/BM/CM/DM Series Monitors

Using CR100 / ColourSpace / BoxIO

Flanders Scientific, Inc.

6215 Shiloh Crossing

Suite G

Alpharetta, GA 30005

Phone: +1.678.835.4934

Fax: +1.678.804.1882

E-mail: [email protected]

www.FlandersScientific.com

Summary of Contents for AM Series

Page 1: ...M CM DM Series Monitors Using CR100 ColourSpace BoxIO Flanders Scientific Inc 6215 Shiloh Crossing Suite G Alpharetta GA 30005 Phone 1 678 835 4934 Fax 1 678 804 1882 E mail Support FlandersScientific com www FlandersScientific com ...

Page 2: ...Management LUT Bypass 3D LUT Start ColourSpace and click the profiling icon From the profiling window select the Settings tab and set Patch Scale to Extended Now select the Hardware Options tab From the Hardware drop down menu select FSI BoxIO Enter the IP address of the BoxIO into the Network Address field then click connect Below the connect button you will see a drop down menu labeled Calibrati...

Page 3: ...We suggest setting extra delay to at least 0 5 sec onds as a setting of 0 seconds will cause read errors Next ensuring that grey is still being displayed on your monitor and the probe is positioned in front of the screen click on Probe Calibration to establish sync for your CR100 and press OK Plug the CR100 into the computer and aim it towards the middle of the monitor Click on the Manual Measure ...

Page 4: ...t supported Then click Create to generate your calibration LUT Once the profile is saved open a LUT Tools window in ColourSpace Once your profile completes click on the graph options tab then select Rename to give the profile an easy to identify name then press save to save the completed profile Click on the Display Characterization icon then select a desired profiling mode Grey Ramp RGB Large qui...

Page 5: ...When the LUT update is complete set LUT Bypass back to None on the monitor Then select your corresponding LUT position User1 User2 User3 from the Color Space Selec tion in the Color Management Display Alignment menu to activate your LUT Next click on the Space Management icon Expand the Working LUTs section and select the LUT that was just saved Select FSI from the drop down menu on the left and 3...

Page 6: ... response For both of these reasons it is advisable to calibrate to a target of 2 4 gamma when calibrating an OLED with black at 0 nits so that noise or junk readings are not being factored into your calibration s EOTF Once calibration is complete you can use the bright knob key on the monitor to adjust your black level higher if desired When calibrated to a target of 2 4 with black at 0 the brigh...

Page 7: ...250 OLED and does not apply to any other models The calibration process remains largely the same as the standard 3D LUT procedure with one important additional step The rest of the calibration procedure then continues per the standard 3D LUT calibration instructions Appendix 2 Calibration of DM250 running Firmware 2 0 0 2282 or later 1 When creating the 3D LUT you will want to set the Max Luminanc...