Original manual – Iss. 0.1
User manual – BM 3E Freezer
12/ 44
Room temperature should not exceed 35°C. The moisture content in the air must not exceed 50% RH. The use of an
air conditioning system can considerably extend the service life of the compressors.
If backup systems using liquid CO
2
or nitrogen vapour injection are used, please refer to the corresponding safety data
sheets in section 13 SAFETY.
5.5
Construction and insulation
The self-contained outside casing is made from zinc electroplated steel sheet protected by epoxy paint.
The inner chamber is made from stainless steel sheet. Thermal insulation is provided by vacuum and polyurethane
foam insulating panels. The insulated swivel door is mounted on a hinge.
The door is closed and sealed by a progressive sealing handle.
5.6
Control console
The control console is located towards the top of the door. It comprises a main screen and a touch-sensitive alarm
window. The main screen is touch-sensitive only on the Evolution model.
Interactive touch-sensitive screen (Evolution
model)
Touch-sensitive alarm window
Lockable handle
Optional bar code reader
Filter cassette
Lower door
Front
view
Opening of lower door panel
Stabilising legs
Alarm warning lights
Disk recorder unit (optional)
Screen (Essential model)
Essential model
Evolution model
Main screen
Touch-sensitive
alarm window
Touch-sensitive
main screen
Touch-sensitive
alarm window