8.3.4 Removing the wet measuring cell
Once the measuring circuit is completely drained, it is possible that some
liquid residues remain in the measuring cell itself. As a precaution use a
paper towel or similar to immediately collect liquid flowing out.
1.
Loosen the eccentric tensioner.
2.
On the opposite side you can now rotate the red twist lock by
180°.
3.
Fold the locking plate open and remove it.
4.
You can take out the flow plate now.
Clean the measuring cell glasses as described in the following chapter.