7.
From this measuring results folder you can select the SOP "Clean /
Fill" already pre-installed in the software, by clicking on "Measure-
ment".
SOP refers to a Standard Operating Procedure
used in our software. In an SOP, work steps are
pre-defined and processed with a single key-
stroke.
Ensure that you select the SOP which is suitable
for your device.
8.
Click on "Start Measurement" to start the flushing or filling proce-
dure.
This procedure removes dirt or the previously analysed
material from the measuring circuit after a measure-
ment.
5.1.4 Sample addition
1.
Start the desired SOP in the "MaS control" program. (see Chapter
Chapter 6.1.1 ‘Fritsch test powder F500’ on page 31)
2.
When requested to add the sample:
n
Add suspensions in small doses to the sample bath using a pip-
ette until the beam absorption set in the SOP is reached.
n
Solids can be put directly into the sample bath using a small
spatula, until the beam absorption set in the SOP is reached.
3.
The measurement starts automatically once the parameters set in
the SOP are reached.
Supply units