Tapping the “MODE” key once more will advance the unit to SCREEN #3.

SCREEN #3 will scroll “Intensity xxxx” to indicate absolute UV measured intensity

2

2  

times 10, in µW/cm  where “xxxx” is the value of measured intensity in µW/cm x 10.

For example:

SCREEN #3: “Intensity 1234” (the absolute measured

2

intensity is 1234 x 10 – 12,340 µW/cm ).

Tapping the “MODE” key once more will advance the unit to SCREEN #4.

SCREEN #4 will display “dxyz” to indicate the number of days elapsed since

the baseline* reset where “d” signifies days and “xyz” is the measured value

of days elapsed.

For example:

SCREEN #4: “d321” (the lamp has been used for 321 days

 since the baseline* reset).

*This is to reset the baseline of all measurements with the exception of the absolute intensity

 measurement (SCREEN #3).

Tapping the “MODE” key once more will advance the unit to SCREEN #5.

SCREEN #5 will scroll “Hold To Reset”. By holding down the “MODE” key for more

than 3 seconds the radiometer will reset the lamp percentage’s reading to 100%

and days elapsed to 0. 

3

Data logging output 

(optional)

Mounting the Sensor

Mounting the Controller

If this option is installed, a linearized analog voltage is output with the following

end points:

The data logging voltage will be set according to the baseline reset procedure

outlined above.

The UV sensor must be mounted facing the lamp(s), approximately 3-10” (8-25 cm)

away from the lamp(s). If the radiometer displays “Sensor Saturated” – the sensor

2

has detected UV intensity greater than 21,300 µW/cm . Move the sensor away from

the lamps until this condition is resolved.

Using two screws, mount the sensor on a at, clean surface where no water will drip

directly on the sensor.

For optimal performance be sure to keep the sensor clean. To clean the sensor, we

recommend using a cotton swab and isopropyl alcohol with compressed air.

 The controller should be mounted outside of the air handling unit.

 Mount in an area with a temperature less than 50˚C and 90% RH.

 Mount on a at, clean surface within the range of the cable.

 Use two or three screws to mount the controller by using the keyholes in the

 back of the controller.

 Connect the sensor to the output port of the controller using the supplied cable

 Connect power to the controller

0% = 0 Volts 

100% = 10 Volts ± 500 mVolts

Note: Using any cable other than the supplied cable may ruin the unit and will void warranty

- 0 mVolts

+ 100 mVolts