Page : 56
NT 0928 Rev.A0
2. Select a 50 ml Braun Perfusor syinge, 20 ml full.
3. Select a flow rate of 120 ml/hr.
q
For normal flow rates, the end of infusion pre-alarm is activated 5 minutes before end of infu-
sion.
q
For higher flow rates (50 ml/hr), the pre-alarm is activated when the volume remaining to be in-
fused equals 10% of total syringe capacity.
4. Ensure the end of infusion pre-alarm is present.
5. Press the SILENCE ALARM key to silence the acoustic alarm and check the end of infusion alarm
(see in 5.4.2.7.2.).
5.10.2. End of infusion alarm
1. At the end of infusion:
q
Check the acoustic and visual end of infusion alarm.
over view
2. Measure the " hard height ": X, with 38.0 < X < 39.0 for a 50 ml Braun Perfusor syringe.
q
Important
:
For accurate checking of the hard height, do not move the driving block when measuring.
3. If X value is out the tolerated ones see "
Eta6
", section 3.2.,CALIBRATION.
5.11. Checking the Led's and keyboard
Refer to
tSt2
tSt2
and
tSt3
tSt3
(see section 3.2.: CALIBRATION)
5.12. Battery autonomy test
1. Medium battery autonomy is 7 hours (minimum 5) when the device operates with a 50 ml Braun
Perfusor syringe at a flow rate of 5 ml/hr (without Master module connection type)
2. The battery discharge pre-alarm warns the user the remaining autonomy is roughly 60 minutes (at
5 ml/h) before infusion will completely stop (total battery discharge alarm).
3. Battery recharging time is 16 hours minimum (100% of its capacity).
5.13. Continuity test
Using an multimeter connected ohmmeter, check the presence of an infinite electrical resistance be-
tween:
q
phase and metal tube,
q
neutral and metal tube.
Summary of Contents for INJECTOMAT 2000 Anasthesie
Page 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Page 2: ......
Page 4: ......
Page 8: ...Page 4 NT 0928 Rev A0 ...
Page 47: ...NT 0928 Rev A0 page 43 ...
Page 53: ...NT 0928 Rev A0 page 49 ...
Page 54: ...Page 50 NT 0928 Rev A0 ...
Page 55: ...NT 0928 Rev A0 page 51 ...
Page 56: ...Page 52 NT 0928 Rev A0 ...
Page 67: ...NT 0928 Rev A0 page 63 ...
Page 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Page 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Page 77: ...NT 0928 Rev A0 page 73 ...
Page 78: ...Page 74 NT 0928 Rev A0 ...
Page 79: ...NT 0928 Rev A0 8 ADDENDA ...
Page 80: ...NT 0928 Rev A0 ...
Page 82: ...NT 0928 Rev A0 ...