Fresenius Medical Care T688 Series Instruction Manual Download Page 3

FRESENIUS MEDICAL CARE       

MEDICAL TREATMENT CHAIR       

INSTRUCTION MANUAL - T688 Series 

Page 2 of 20

 

 

 

IFU T688 EN V7 Apr 2021

 

 

 

 

INDICATIONS FOR USE & GENERAL SAFETY  

1.1 

Indications for Use 

 

 

The Medical Treatment Chair, T688 Series chair is intended for use in medical procedures such 

as  the  administration  of  renal  dialysis  to,  and  collecting  blood  from,  patients  in  hospital 
departments, under the supervision of trained medical staff. The Medical Treatment Chair is also 
intended  for  use  in  day  surgery.  The  T688  Series  chair  is  designed  so  that  the  occupant  is 
accommodated  in  a  seated  position  with  the  hips  moved  back  so 

that  the  occupant’s  back  is 

against  the  back  rest  and  the  legs  outstretched  and  supported  by  the  seat  and  leg  rests.  The 
chair is designed in such a way that no special training nor competence is needed for operating it. 

The T688 Series chair is also used to position patients for easy access by healthcare 
professionals. 

The T688 Series chair is intended to be used by patients with a weight not exceeding 200 kg (440 
lbs). 

1.2 

General Safety 

 

The T688 Series chair shall only be used indoors on flat surfaces under the supervision of 
a medical staff 

The chair should never be used in a manner for which it was not intended. 

The chair is designed so that the occupant will be seated in a typical seated position - hips moved 
back  so  that  the  occu

pant’s spine is against the backrest, with legs outstretched and supported 

by the seat and leg rest.  

The  chair  is  fitted  with  castors,  which  are  specifically  used  to  aid  cleaning  and/or  positioning  of 
the chair  within the room.  

The T688  Series chair  shall not be used to transport patients or 

any other items

.  

Ensure that the backrest is fully elevated before moving the chair. 

For models fitted with dual hand controls: the programmable unit (

Nurse’s Hand Control) is for the 

use of the staff members only, and the dual recline unit (

Patient’s Hand Control), where fitted, is 

to  be  used  by  the  patient.   

This  ensures  that  the  chair  is  not  raised  or  lowered,  by  the 

occupant.  A  staff  member  should  ensure  the  area  surrounding  the  chair  is  free  from 
obstruction  before  adjusting  it.  The  Nur

se’s Hand Control should be kept out of reach of 

the patient at the rear of the chair. 

It is the staff  member's responsibility to  ensure that prior to operating  the chair,  the surrounding 
walls or equipment will not come into contact with nor obstruct the free movement of the chair. 

It is the staff members' responsibility to ensure the patient is briefed on the safe operation 
of  the  chair.    If  the  staff member  believes  the  patient  cannot  safely  operate  the  chair,  the 
staff member should remove the hand cont

rol from the patient’s reach.  

In  an  emergency,  the  chair  can  be  lowered  to  the  Trendelenburg  position  where  CPR  may  be 
performed. It is the staff members' responsibility to ensure that the chair is in the correct position 
for CPR and that the resuscitation method used 

is in line with hospital policy.

  In addition, staff 

should  ensure  that  the  chair  is  adequate  for  those  resuscitation  methods, 

in  line  with  hospital 

policy. Do not sit on the backrest, arms, or the leg rest of the chair. 

The upholstery is applied parts. The upholstery has no electrical connection to other parts of the 
chairs, and is made of non-conducting material. 

In general, no responsibility or liability can be accepted by the manufacturer for failure to adhere 
to the guidelines and instructions contained in this manual. 

Summary of Contents for T688 Series

Page 1: ...Medical Treatment Chair T688 Series Instruction Manual ...

Page 2: ...ROLS PROGRAMMING 7 3 3 1 Hand Control and optional Seat Height Foot Control 7 3 3 2 Programming Instructions 9 3 3 3 Recalibration 9 4 MAINTENANCE 10 4 1 INSPECTION SERVICING 10 4 2 CLEANING 10 4 3 FABRIC STANDARDS 11 4 4 TROUBLESHOOTING 11 5 ARRANGING A SERVICE 12 6 WARRANTY INFORMATION 13 7 WASTE DISPOSAL 13 8 PHYSICAL DESCRIPTION 14 8 1 DESCRIPTION OF DEVICE 14 8 2 TECHNICAL SPECIFICATIONS 14 8...

Page 3: ...be used to transport patients or any other items Ensure that the backrest is fully elevated before moving the chair For models fitted with dual hand controls the programmable unit Nurse s Hand Control is for the use of the staff members only and the dual recline unit Patient s Hand Control where fitted is to be used by the patient This ensures that the chair is not raised or lowered by the occupan...

Page 4: ... Pressure sores 1 5 Reporting Serious Incident Any serious incident that has occurred in relation to this medical device shall be reported to the manufacturer and local health authority in accordance with the local regulations Within the EU the user must report any serious incident that has occurred in relation to the device to the manufacturer according to the labelling and the competent authorit...

Page 5: ...per instruction 5 above 9 Arms maybe swung out or lowered flat depending on style to seat level to facilitate side transfer of patient Raise or lower the seat height to prevent unnecessary bending or lifting by staff members 10 IV Pole if fitted can be removed or repositioned by loosening the knob at the lower end of the mounting bracket and retightening the knob 11 Lift up tray if fitted can be r...

Page 6: ...the nurse hand control or the patient hand control if fitted 20 Seat Height Foot HI Lo Control if fitted can be used raise or lower chair height 21 USB charger if fitted can be used to charge a USB device such as a mobile phone a music player or a camera Maximum output of the charger is 4 watt NOTE It may not charge all USB devices depending on the power input requirement of the USB devices NOTE T...

Page 7: ... point should be switched on Refer to section 3 2 The backup batteries if fitted will recharge automatically and slow to trickle charge when full It is important to charge the backup battery for a minimum of 24 hours before use the chair 3 1 WARNINGS AND PRECAUTIONS The following table shows potential hazards that have been identified and the steps that should be taken to avoid them IDENTIFIED HAZ...

Page 8: ...ntative at the details provided in section 5 The battery is not user replaceable The chair must be connected to the power supply to recharge at least every 5 days 3 3 HAND CONTROLS PROGRAMMING Each T688 Series chair can be supplied with two hand controls One nurse operated hand control and one patient operated hand control The patient s individual hand control allows operation of the seat footrest...

Page 9: ...FRESENIUS MEDICAL CARE MEDICAL TREATMENT CHAIR INSTRUCTION MANUAL T688 Series Page 8 of 20 IFU T688 EN V7 Apr 2021 Seat Height Foot Control ...

Page 10: ...ive seconds NOTE Both buttons must be activated at exactly the same time If beeping does not start repeat until successful Stage 2 Initialise the System Process 1 Press both of the Up and Down arrow buttons for the Backrest function simultaneously and hold until all the actuators have stopped moving This will drive the Backrest actuator into the fully retracted down the Leg Rest actuator into the ...

Page 11: ...faulty and further damage occurs 4 2 CLEANING Cleaning should conform to the standards set by the hospital All materials used in the manufacture of the chairs have been found to be suitable to be cleaned with standard hospital cleaning products however to ensure suitability a test should be carried out on an inconspicuous piece of material or a sample provided by the manufacturer The following sho...

Page 12: ... Check that the chair is connected into the mains power outlet and switched on correctly as indicated by the green indicator light on the nurse s hand control 3 Ensure that the batteries have been on charge for 24 hours minimum This should occur at least every 5 days 4 Ensure there are no damage or cuts in the cables to the hand control or actuator cables 5 Determine functions are not operating as...

Page 13: ...Office Note Please quote the chair serial number which can be found on the Type Label attached to the chair An example of the Type Label is in section 8 2 7 CONTACT INFORMATION https fmc seating com MANUFACTURER FRESENIUS MEDICAL CARE SEATING AUSTRALIA PTY LTD 786 Stud Road Scoresby VICTORIA 3179 AUSTRALIA sales seatingAU fmc asia com SERVICE TEL 61 0 3 9780 9509 SALES TEL 61 0 3 9780 9500 FAX 61 ...

Page 14: ...2 month warranty 6 T688 Series Medical Treatment Chairs are not covered by warranty if used outdoors 7 The use of unauthorised labour will void the warranty Contact Fresenius Medical Care for approval prior to commencing work 8 If you are unsure about any aspect of this warranty please contact Fresenius Medical Care at the number given in Section 5 of this manual If you need to arrange a service p...

Page 15: ...0 cm Width 100 cm Length 99 cm Weight depending on options 105kg Capacity 200 kg 8 2 2 Electrical Safety Classification according to IEC 60601 1 2012 Degree of protection against electric shock Type B CO7 Control Box Specifications Degree of protection against ingress of solids liquids IPX6 Mains voltage 100 240V 50 60 Hz Battery Capacity 1 3 Ah 24 VDC Hand Control Specifications Degree of protect...

Page 16: ...021 8 2 4 Operating Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 5 Transport Storage Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 6 Electromagnetic or other interference Switch off devices which can cause electromagnetic interference 8 2 7 Type Label ...

Page 17: ...on 2 Memory Position 3 Stores memory position Hand Control Indication Lights LED Amber Off and LED Green On indicates Mains is on and Battery is fully charged LED Amber On and LED Green Off indicates Battery is in normal use LED Amber blinking and LED Green Off indicates Battery is in normal use and requires charging LED Amber ON and LED Green ON indicates mains is on and battery is charging LED A...

Page 18: ...age 17 of 20 IFU T688 EN V7 Apr 2021 pressed Seat Height Foot Control optional Symbol Raise the Chair Lower the Chair Authorised Representative in the European Community Fresenius Medical Care Deutschland GmbH Else Kröner Straße 1 61352 Bad Homburg Germany Telephone 49 6172 609 0 ...

Page 19: ...mpliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply l...

Page 20: ... in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propag...

Page 21: ...to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 3 691 1 17 1 17 2 33 1011 67 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommen...

Page 22: ...IFU T688 EN V7 April 2021 ...

Reviews: