Fresenius Medical Care T688 Series Instruction Manual Download Page 1

 

 

Medical Treatment Chair

 

 

T688 Series

 

Instruction Manual

 

 
 
 

 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for T688 Series

Page 1: ...Medical Treatment Chair T688 Series Instruction Manual ...

Page 2: ...ROLS PROGRAMMING 7 3 3 1 Hand Control and optional Seat Height Foot Control 7 3 3 2 Programming Instructions 9 3 3 3 Recalibration 9 4 MAINTENANCE 10 4 1 INSPECTION SERVICING 10 4 2 CLEANING 10 4 3 FABRIC STANDARDS 11 4 4 TROUBLESHOOTING 11 5 ARRANGING A SERVICE 12 6 WARRANTY INFORMATION 13 7 WASTE DISPOSAL 13 8 PHYSICAL DESCRIPTION 14 8 1 DESCRIPTION OF DEVICE 14 8 2 TECHNICAL SPECIFICATIONS 14 8...

Page 3: ...be used to transport patients or any other items Ensure that the backrest is fully elevated before moving the chair For models fitted with dual hand controls the programmable unit Nurse s Hand Control is for the use of the staff members only and the dual recline unit Patient s Hand Control where fitted is to be used by the patient This ensures that the chair is not raised or lowered by the occupan...

Page 4: ... Pressure sores 1 5 Reporting Serious Incident Any serious incident that has occurred in relation to this medical device shall be reported to the manufacturer and local health authority in accordance with the local regulations Within the EU the user must report any serious incident that has occurred in relation to the device to the manufacturer according to the labelling and the competent authorit...

Page 5: ...per instruction 5 above 9 Arms maybe swung out or lowered flat depending on style to seat level to facilitate side transfer of patient Raise or lower the seat height to prevent unnecessary bending or lifting by staff members 10 IV Pole if fitted can be removed or repositioned by loosening the knob at the lower end of the mounting bracket and retightening the knob 11 Lift up tray if fitted can be r...

Page 6: ...the nurse hand control or the patient hand control if fitted 20 Seat Height Foot HI Lo Control if fitted can be used raise or lower chair height 21 USB charger if fitted can be used to charge a USB device such as a mobile phone a music player or a camera Maximum output of the charger is 4 watt NOTE It may not charge all USB devices depending on the power input requirement of the USB devices NOTE T...

Page 7: ... point should be switched on Refer to section 3 2 The backup batteries if fitted will recharge automatically and slow to trickle charge when full It is important to charge the backup battery for a minimum of 24 hours before use the chair 3 1 WARNINGS AND PRECAUTIONS The following table shows potential hazards that have been identified and the steps that should be taken to avoid them IDENTIFIED HAZ...

Page 8: ...ntative at the details provided in section 5 The battery is not user replaceable The chair must be connected to the power supply to recharge at least every 5 days 3 3 HAND CONTROLS PROGRAMMING Each T688 Series chair can be supplied with two hand controls One nurse operated hand control and one patient operated hand control The patient s individual hand control allows operation of the seat footrest...

Page 9: ...FRESENIUS MEDICAL CARE MEDICAL TREATMENT CHAIR INSTRUCTION MANUAL T688 Series Page 8 of 20 IFU T688 EN V7 Apr 2021 Seat Height Foot Control ...

Page 10: ...ive seconds NOTE Both buttons must be activated at exactly the same time If beeping does not start repeat until successful Stage 2 Initialise the System Process 1 Press both of the Up and Down arrow buttons for the Backrest function simultaneously and hold until all the actuators have stopped moving This will drive the Backrest actuator into the fully retracted down the Leg Rest actuator into the ...

Page 11: ...faulty and further damage occurs 4 2 CLEANING Cleaning should conform to the standards set by the hospital All materials used in the manufacture of the chairs have been found to be suitable to be cleaned with standard hospital cleaning products however to ensure suitability a test should be carried out on an inconspicuous piece of material or a sample provided by the manufacturer The following sho...

Page 12: ... Check that the chair is connected into the mains power outlet and switched on correctly as indicated by the green indicator light on the nurse s hand control 3 Ensure that the batteries have been on charge for 24 hours minimum This should occur at least every 5 days 4 Ensure there are no damage or cuts in the cables to the hand control or actuator cables 5 Determine functions are not operating as...

Page 13: ...Office Note Please quote the chair serial number which can be found on the Type Label attached to the chair An example of the Type Label is in section 8 2 7 CONTACT INFORMATION https fmc seating com MANUFACTURER FRESENIUS MEDICAL CARE SEATING AUSTRALIA PTY LTD 786 Stud Road Scoresby VICTORIA 3179 AUSTRALIA sales seatingAU fmc asia com SERVICE TEL 61 0 3 9780 9509 SALES TEL 61 0 3 9780 9500 FAX 61 ...

Page 14: ...2 month warranty 6 T688 Series Medical Treatment Chairs are not covered by warranty if used outdoors 7 The use of unauthorised labour will void the warranty Contact Fresenius Medical Care for approval prior to commencing work 8 If you are unsure about any aspect of this warranty please contact Fresenius Medical Care at the number given in Section 5 of this manual If you need to arrange a service p...

Page 15: ...0 cm Width 100 cm Length 99 cm Weight depending on options 105kg Capacity 200 kg 8 2 2 Electrical Safety Classification according to IEC 60601 1 2012 Degree of protection against electric shock Type B CO7 Control Box Specifications Degree of protection against ingress of solids liquids IPX6 Mains voltage 100 240V 50 60 Hz Battery Capacity 1 3 Ah 24 VDC Hand Control Specifications Degree of protect...

Page 16: ...021 8 2 4 Operating Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 5 Transport Storage Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 6 Electromagnetic or other interference Switch off devices which can cause electromagnetic interference 8 2 7 Type Label ...

Page 17: ...on 2 Memory Position 3 Stores memory position Hand Control Indication Lights LED Amber Off and LED Green On indicates Mains is on and Battery is fully charged LED Amber On and LED Green Off indicates Battery is in normal use LED Amber blinking and LED Green Off indicates Battery is in normal use and requires charging LED Amber ON and LED Green ON indicates mains is on and battery is charging LED A...

Page 18: ...age 17 of 20 IFU T688 EN V7 Apr 2021 pressed Seat Height Foot Control optional Symbol Raise the Chair Lower the Chair Authorised Representative in the European Community Fresenius Medical Care Deutschland GmbH Else Kröner Straße 1 61352 Bad Homburg Germany Telephone 49 6172 609 0 ...

Page 19: ...mpliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply l...

Page 20: ... in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propag...

Page 21: ...to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 3 691 1 17 1 17 2 33 1011 67 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommen...

Page 22: ...IFU T688 EN V7 April 2021 ...