FreiStil Tischlerei LUKAS 136 Operating Instructions Manual Download Page 8

 

 

 

BT 2018 Lukas 136/170 – 01  

 

 

 

 

 

 

 

Version – 05/2018

 

 

- 8 - 

2. OPERATING INSTRUCTIONS 

2.1 

Intended use

 

The 

“LUKAS” 

children’s therapeutic bed is designed for use in a 

home-care environment. 
Application environment 3 + 4 in accordance with DIN EN 60601-2-52. The medical treatment to be 
provided is the diagnosis, treatment and monitoring of illnesses under medical supervision. It is 
permitted for use for 

one person only

 

Height

: min. 75cm – max. bed base minus 15cm/

weight

: max. 135 kg - 

and the safe working load (see type plate) must not be exceeded. The bed must only be used in 
enclosed rooms and must be protected against excessive moisture. 
The user can apply the different adjustment functions using an 

IPX4 manual control unit with closure, 

height and bed base adjustment

Make sure that the patient is always lying with their head at the “head end” of the bed. 
All components of the bed must be used in accordance with the appropriate intended use. Overloading causes damage 
and is not permitted. 
 
 

2.1.1 Indications 

For children with restricted movement who no longer have spontaneous mobility, or young people and 
adults with cognitive developmental disorders, for example it  

 

An ergonomic sitting position is necessary in bed due to trunk instability  

 

Secure contact between the feet and the floor is necessary when getting into/out of bed 

 

The height of the bed must be adjustable so as to enable the patient to be transferred to a 
wheelchair, for example 

 

The height of the bed base must be adjustable for treatment purposes 

2.1.2 Contraindications

 

 

None known 

Always consult a doctor or therapist before using the bed. 

2.2  Important information on operational safety

 

I. 

Setup and initial operation are performed by the sales agent. Prior to handing over the bed and 
before each re-use, a functional check and a safety check must be performed. It must be ensured 
that all functions are running smoothly, the power cable is routed correctly, and the bed base is 
securely mounted in the bed frame. This also applies to any further operation of the children’s 
therapeutic bed. 
Incorrect routing of the power cable (e.g. kinks, shearing) may pose a serious risk to people 
(electric shock). 

II. 

The patient can weigh up to a maximum of 135 kg. Persons of any age with reduced perception 
are only permitted to use the system under supervision. 

 
 
 
 
 

Summary of Contents for LUKAS 136

Page 1: ...Phone 49 0 541 800 3969 0 Fax 49 0 541 800 3969 99 Email info freistil com Website www freistil com OPERATING INSTRUCTIONS Children s Therapeutic Bed LUKAS door height 136cm door height 170cm Read ca...

Page 2: ...ng instructions 2 1 Intended use 07 2 2 Important information on operational safety 07 2 3 Warning notices 08 2 4 Lifting unit 09 2 5 Bed base 09 2 6 Manual control unit 10 2 7 Revolving doors 11 2 8...

Page 3: ...here are stickers with arrows numbered from 1 to 4 Fig 1 Arrange the side panels the head end and the foot end according to these numbers Fig 1 Screw the side panels to the head end and foot end accor...

Page 4: ...sition the lifting scissors according to the head end sticker in the bed frame Fig 3 Move the lifting scissors to the uppermost position Fig 4 Screw the lifting scissors onto the supporting strip usin...

Page 5: ...rs adapter using the enclosed flat head screws 4 x 35 mm Fig 8 Fig 7 Fig 8 If present connect the cable for the dual motor to the control unit Fig 9 The control unit is located on the lifting scissors...

Page 6: ...d end foot end of the bed frame according to the colour coding using the enclosed screws M8 x 100 mm Fig 13 Then attach the second door unit accordingly Fig 13 Insert or remove spacers betwen the door...

Page 7: ...them opening unintentionally Fig 19 Fig 19 1 8 Adjusting the doors It is possible to correct stiff or hanging doors using a size 3 Allen key by unscrewing the support foot in the outer door hinges Tog...

Page 8: ...ntaneous mobility or young people and adults with cognitive developmental disorders for example it An ergonomic sitting position is necessary in bed due to trunk instability Secure contact between the...

Page 9: ...structions item 2 3 2 4 and type plate VII In the event of misuse improper use incorrect operation or incorrect assembly repair no liability will be accepted for any damage In such cases all warrantie...

Page 10: ...the head end or the head end foot end accessories must only be used to lift and lower the upper body and legs observe the position of the patient in the bed 2 5 Bed base The lifting unit is used for t...

Page 11: ...nabling and disabling the function keys is located on the back of the manual control unit Fig 22 Turn the key to enable and disable the manual control unit The arrow indicates the relevant position Th...

Page 12: ...ed wooden strip Fig 25 In this case the doors are locked by turning the locking pins Fig 23 Fig 24 Fig 25 2 8 Transport wheels Caster 75 x 30mm To secure the lock press the foot lever down To release...

Page 13: ...intenance safety checks and inspection intervals see item 4 of the ANNEX to the operating instructions Spare parts lists can be obtained from the manufacturer 2 12 Accessories Mattress not readily fla...

Page 14: ...0 x 200 cm External dimensions 1072 x 2160 mm Bed base 1000 x 2000 mm Mattress size 1000 x 2000 mm Adjustment range of the lifting scissors 390 830 mm Bed height wheel diameter 75 mm 2260 mm Door heig...

Page 15: ...ckrest rotary points of back section seating section upper lower leg section 94 Weight approx 20 kg User load 135 kg Safe working load 170 kg 3 4 Motors Limoss Mains voltage frequency 230 volts 50 Hz...

Page 16: ...Inspections performed as part of the scheduled maintenance work can be performed on the installed system at site of installation All other repairs must be performed at a suitable place of work Only t...

Page 17: ...e wheels of the bed are working correctly The moving and braking functions must be working correctly For electrically adjustable beds perform a visual and functional check of the drives These are subj...

Page 18: ...136 170 01 Version 05 2018 18 Maintenance safety check in accordance with 6 of the German Medical Device Operator Ordinance MPBetreibV This form is available to download from our website www freistil...

Reviews: