999075 – Revision B
P a g e
|
34 of 48
Weights
Safe Working Load (SWL)
130kg / 160kg / 200kg /270kg
Hoist
10 kg (22 lb)
Battery charger
0.8kg (1.76 lb)
Carry bar
2 kg (4.4 lb)
Handset
0.2kg (0.44 lb)
Operational Forces
Handset
3N
Emergency cord
15N
Spring clips on carry bar
8N
Manually traversing fully loaded hoist (SWL)
50N
Manually traversing unloaded hoist (No weight)
10N
Expected Product Lifetime
Ten years depending usage and compliance to maintenance, servicing and LOLER inspections.
Serviceable parts within this period are batteries and the lift tape. Batteries should have an expected service life of 200
discharge cycles or 3 years, dependant on the charging routine. The lift tape should have an expected service life of 2 years if
used correctly but visual inspection should be carried out before use.
Standards Applied
The standards that have been applied to the device are as follows:
•
EN 10535:2006 Hoists for the transfer of disabled persons. Requirement and test methods.
•
EN 60601-1-1:2006 +A12:2014 Medical electrical equipment. General requirements for basic safety and essential
performance.
•
EN 60601-1-2:2015 Medical electrical equipment. General requirements for basic safety and essential performance.
Collateral Standard. Electromagnetic disturbances. Requirements and tests.
•
EN 60601-1-6:2010 Medical electrical equipment. General requirements for basic safety and essential performance.
•
EN 60601-1-11:2015 Medical electrical equipment. General requirements for basic safety and essential performance.
Table 12
Table 13