background image

 

 

Geiger counter Handheld w/ built-in tube   513545 

 
 

 

2020-02-27 / MRA 

Product manual

 AC513545

   p.

 3/4

 

Dansk 

manual: 

Se p. 1

  

 

 

QUICK GUIDE 

Power on:

 Press ON to power on the instrument. 

The  display  briefly  shows  a  version  number  and 
the GM voltage. Immediately after the counter has 
been turned on, the counting period is 1 s 

 

CPS

 is 

indicated on the display. 

Other  views:

  From  CPS  press 

  once  to  select 

µSv/h 

relative to Cs-137. 

Press 

  once  more  to  select  the  mean  value  of 

counts per second 

 

Mean CPS

Other counting periods:

 Press 

 or 

 until 

Pulse

 

is highlighted in the display. The instrument now 
counts for a selectable period, starting with 10 s. 

Press  HOLD 

  HOLD  and  set  the  measurement 

period with 

 and 

The  unit  is  seconds.  The  symbol  “

---

“means 

indefinite counting. 
Select with 

CAL

 / cancel with 

HOLD

Setting the GM voltage:

 Press 

CAL

 and adjust the 

voltage with

 and 

.  Select with 

CAL

.  

Counting  rate  (

CPS

)  is  shown  in  the  secondary 

display during the adjustment. 
This setting returns to the default value when the 
instrument is turned on. 

 

8 fixed count periods (1, 10, 20, 30, 40, 50, 
60 or 100 s) 

 

Display of dose rate (µSv/h) related to Cs-
137 

 

GM tube voltage can be adjusted between 
400 V and 700 V while showing count rate. 

 

Max, Min and average view for µSv/h and 
CPS. 

Description 

The GM tube 

The Geiger tube is provided with a cap to protect 
the sensible window. When working with gamma 
radiation,  the  cap  may  remain  in  place.  Remove 
the cap when alpha or beta radiation is involved. 

NOTE:

 Although the GM tube itself is protected by 

a plastic housing, the tube itself is relatively fragile 
in its construction. Particular attention should be 
paid to the radiation stop of the sensor when the 
cap is removed in connection with Alpha and Beta 
measurements. Therefore, do not drop or squeeze 
the  instrument  or  touch  /  strike  the  radiation 
peak. 

If the accident is out as described above, the pipe 
will typically crack. However, it is possible to have 
the tube replaced - A service assignment we can 
offer at an agreed price. 

Time 

The  counting  period  1  s  is  achieved  by  selecting 

CPS

  (counts  per  second).  All  other  periods, 

including  infinity  is  obtained  with  the 

Pulse

 

setting, When the instrument has been turned off, 
the measurement period for 

Pulse

 is 10 s. 

To  change  the  period,  press 

HOLD

 

 

HOLD

  and 

set the measurement period with 

 and 

The unit is seconds. The symbol “

---

“means infinite 

counting. 
Select with 

CAL

 / cancel with 

HOLD

During Pulse measurements, the progress of each 
measurement  can be  followed  on the %  scale  at 
the top 