21
4.0 Startup Overview continued
4.1 Parameter settings to check for systems with Fluoride modules only
Check by entering “setup” mode and ensure parameter settings match the defaults shown on page
6. To change any parameter you must first turn off the software lock, parameter P01.
i)
For systems supplied with ISE measurement module (transmitter) only
:
(1) Make sure the operating range (P09) matches the needs of your process: Low (0-9.99
ppm); Mid (0-99.9) or High (0-999)
(2) Make sure the mA input of an external device you connect to matches the scaling of the
transmitter's mA output (P10 & P11).
(3) Make sure the temperature input (P03) is set to “999” for the PT1000 temperature
element.
4.2 Settings to check for systems with alarm relay modules
ii)
For systems supplied with the FX-300-REL alarm relay & control module:
(1) Make sure the analog input (P03) on the REL module matches the analog output of the
ISE module (P08). Selections are 0-20mA or 4-20 mA.
(2) Make sure display mode (P06) on the REL module matches the output variable (P07) on
the ISE module (factory set to ISE).
(3)
Make sure the ISE Output (operating) range (P09) of the ISE
module
matches the ISE
input range (P07) of the REL module.
(4)
Enter the value for the 0/4mA input (P09) on the REL module to match the 0/4 mA output
(P10) of the ISE module.
(5)
Enter the value for the 20mA input (P09) on the REL module to match the 20 mA output
(P11) of the ISE module.
(6)
On the REL module set the type of limit for alarm limit 1: Hi (max), Lo (min). Default is Hi.
(7)
On the REL module set the type of limit for alarm limit 2: off, Hi (max), Lo (min). Default is
Lo
(8) On the REL module set the alarm trip point for each limit by using the limit buttons on the
face of the module. First unlock the software. Return to run (input) mode, press the Mode
button until Limit 1 LED lights, and use the up/down keys until the set point is displayed.
Press Mode button to light the Limit 2 LED, and use the up/down keys to enter the set
point. Press the Mode button to return to run mode. The software will lock automatically
after 10 minutes.
4 Start Up