Change Control
Change Control (initiated with version 1.01):
Version Date
Summary of Changes
1.01
November 16,
2006
Address changes to reflect move to 7703 Montpelier Road, Laurel, MD
20723. This was applied on the cover page and on page 66.
1.02
August 24,
2009
Updated headers and Table of Contents to current style.
Updated Network Connection to current recommendations.
Replaced photos showing XC-1200 Standard on probe, clarified
Reference Standardization explanation.
Updated various menus to current XDS style.
Moved Wavelength Linearization before Reference Standardization.
Replaced section on XDS Instrument Connection with an updated,
more current version. This also includes basic connection
troubleshooting. The sections on Windows 95, 98, and NT 4.0 were
dropped.
Updated explanations for Instrument Calibration and Wavelength
Certification.
Added photos to some procedures to clarify required probe positions.
Added note that USP dropped “Photometric Linearity Test” as of 12-
2008.
Added notes about possible failure of Photometric Test, and corrective
action.
Corrected wrong instruction in Gain Test. Reflectance probe is normally
used in Gain.
Changed photo in Lamp Replacement, now showing arrow on lamp.
Corrected reference to Lamp Replacement in Troubleshooting to
section 8.3.
Updated Vision “opening screen”, removed copyright date for
publication.
Updated Configure, Options screen and discussion to current version.
Added note to Troubleshooting, Instrument Calibration, that Reference
Standardization must be performed.
Added note about wavelength drift due to temperature change with
the WSR standard. This is on the page immediately following
Instrument Calibration.
Added note about 21 CFR Part 11 Verification document to Validation
section.
Added note to avoid scratching sample probe window to maintenance
section.
Added troubleshooting items which may be symptomatic of network
connection issues.
Added note on page in section on Interactance Reflectance, to avoid
carrying dirt or sample material from the OptiProbe barrel to the
instrument reference.
1.03
February 21,
2012
Corrected photo and description of taking reference during
Wavelength Linearization.