4. BATTERY CHARGING 

The ANALYZER has a built-in lithium battery and can be charged via mini-USB 
port. USB Charger must be rated >1.0A for fast charging. Before charging, 
TURN OFF the analyzer to avoid any potential damage. Charging takes about 4 
hours. Do not charge the device in a combustible area. The analyzer has the 
battery level icon to provide you with an indication of how much battery is left. 

 

5. OPERATIONAL TIPS 

ü

 

Before turning the detector off don’t forget to flush/purge out the detector 
by allowing the pump to run for 60 secs to allow clean air to purge. 

ü

 

Ensure tubing and items are clean before storing as to avoid any 
contamination, residual odors or toxic gases that may poison the sensor. 

ü

 

Ensure periodic calibration every 6 months so that the performance of the 
detector remains within specification. If the calibration period is >6 months, 
the detector still operates but accuracy will be compromised. 
 
 

** WARNING** 

 

Ø

 

KEEP DETECTOR

 AWAY

 FROM ELECTROMAGNETC & MAGNETIC INTERFERENCES (i.e. 

PHONES & MAGNETS) 

Ø

 

STORE DETECTOR 

WITHIN SPECIFICATIONS

 

Ø

 

IF UNWELL, SEEK CLEAN AIR & MEDICAL ATTENTION. 

Ø

 

FOLLOW INSTRUCTIONS AS THE DETECTOR IS 

VERY SENSITIVE

 

Ø

 

TO ENSURE

 ACCURACY, CALIBRATE DEVICE AT LEAST 

EVERY 6 MONTHS

 

Ø

 

THE DETECTOR IS NOT RECOMMENDED FOR USE IN FLUE GAS AS THE UNIT DOES NOT 
HAVE A WATER TRAP NOR A NOX FILTER and PROBE WILL NOT WITHSTAND TYPICAL 
FLUE GAS TEMPERATURES. 

Ø

 

CHECK AND SET ALARM LEVELS APPROPRIATELY TO AVOID HARMFUL EXPOSURE - 
CONSULT WITH YOUR SAFETY OFFICER OR WITH STATE/FEDERAL AGENCIES. 

 

 

 

 

Summary of Contents for FD-311

Page 1: ...sicsdetectors com 1 INDUSTRIAL SPEC GAS ANALYZER with BUILT IN PUMP Model FD 311 Manual Print Version 1 0 USER MANUAL Checklist Serial He sensor Ar sensor Other Telescopic Probe Kit 1ft Probe 2 Probe...

Page 2: ...FORENSICS DETECTORSTM FORENSICS DETECTORSTM www forensicsdetectors com 2 GAS SENSOR RANGES DEFAULT ALARM SET POINTS Target Gas Range Resolution Low Alarm High Alarm Helium He 0 100 vol 0 1 vol 20 50 A...

Page 3: ...0 F 122 F Store Oper Humidity 95 RH Battery DC3 7V Li Ion battery 2000mAh Charging Time 4 hours Operation Time 6 hours Dimension Weight 9 5 x 2 5 x 1 7 inches 440grams Rating IP65 Explosion Proof Ex i...

Page 4: ...use for engineering professionals technicians contractors energy inspectors researchers and labs 2 QUICK SETUP 1 Ensure all components are included 2 Ensure analyzer is fully charged If not please ch...

Page 5: ...Concentration You will hear the pump and you are ready to start taking measurements To turn OFF press MIDDLE button for 3 secs Press the left button to Turn off on the PUMP Press the right button to...

Page 6: ...nts are met exceeded To change the alarm set points select AlarmSet in the Main Menu see prior section You can change the preset alarm level to your desired level by using the left and right buttons A...

Page 7: ...com 7 ALARM LOGGING AlarmLog When alarms are triggered they are automatically recorded with a date and time stamp along with the min or max alarm data point This is useful for safety personnel to kee...

Page 8: ...value Select ZeroSet from the Main Menu Screen Expose to ZERO air for 1 min Use fresh air Maintain a flow of 0 5L min to deliver the gas to the detector i e via the inlet Use a T piece Excess gas shou...

Page 9: ...and highly specific applications calibration can be undertaken more often to ensure the highest accuracy If left beyond 6 months the accuracy of the analyzer will be compromised To undertake span cal...

Page 10: ...tectors com 10 MENU SCREEN Check Performs a self check of the analyzer Info Provides the software version information TimeSet Allows to set the date and time Restore Restores to factory default settin...

Page 11: ...n before storing as to avoid any contamination residual odors or toxic gases that may poison the sensor Ensure periodic calibration every 6 months so that the performance of the detector remains withi...

Page 12: ...ranty by third parties We shall not be liable for any incidental or consequential damages caused by the breach of any express or implied warranty Except to the extent prohibited by applicable law any...

Reviews: