INDUSTRIAL GAS ANALYZER FD-311
FORENSICS DETECTORS
TM
FORENSICS DETECTORS
TM
www.forensicsdetectors.com
11
4. BATTERY CHARGING
The ANALYZER has a built-in lithium battery and can be charged via mini-USB
port. USB Charger must be rated >1.0A for fast charging. Before charging,
TURN OFF the analyzer to avoid any potential damage. Charging takes about 4
hours. Do not charge the device in a combustible area. The analyzer has the
battery level icon to provide you with an indication of how much battery is left.
5. OPERATIONAL TIPS
ü
Before turning the detector off don’t forget to flush/purge out the detector
by allowing the pump to run for 60 secs to allow clean air to purge.
ü
Ensure tubing and items are clean before storing as to avoid any
contamination, residual odors or toxic gases that may poison the sensor.
ü
Ensure periodic calibration every 6 months so that the performance of the
detector remains within specification. If the calibration period is >6 months,
the detector still operates but accuracy will be compromised.
** WARNING**
Ø
KEEP DETECTOR
AWAY
FROM ELECTROMAGNETC & MAGNETIC INTERFERENCES (i.e.
PHONES & MAGNETS)
Ø
STORE DETECTOR
WITHIN SPECIFICATIONS
Ø
IF UNWELL, SEEK CLEAN AIR & MEDICAL ATTENTION.
Ø
FOLLOW INSTRUCTIONS AS THE DETECTOR IS
VERY SENSITIVE
Ø
TO ENSURE
ACCURACY, CALIBRATE DEVICE AT LEAST
EVERY 6 MONTHS
Ø
THE DETECTOR IS NOT RECOMMENDED FOR USE IN FLUE GAS AS THE UNIT DOES NOT
HAVE A WATER TRAP NOR A NOX FILTER and PROBE WILL NOT WITHSTAND TYPICAL
FLUE GAS TEMPERATURES.
Ø
CHECK AND SET ALARM LEVELS APPROPRIATELY TO AVOID HARMFUL EXPOSURE -
CONSULT WITH YOUR SAFETY OFFICER OR WITH STATE/FEDERAL AGENCIES.