Instruction Manual
Aquacon 4000
24
6.3
Setting temperature (Set °C°F) sub-function
SETUP
SETUP
SETUP
HOLD
HOLD
HO LD
4
4
4
ENT
SETUP
SETUP
HOLD
HOLD
4
4
ENT
°C
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
4
4
4
ENT
ENT
ENT
ENT
ENT
ATC
ATC
ATC
°C
°F
1.
Select the “SET °C°F” sub-function, then
press the ENT key
.
2.
Selecting temperature unit: press the
▲
or
▼
key
to select the desired temperature
unit “°C” or “°F”. Press the ENTER key to confirm your selection.
3.
Enable/disable ATC mode: press the
▲
or
▼
key
to enable (ATC ON) or disable (ATC
OFF) automatic temperature compensation.
Press the ENT key
to confirm your selection.
−
ATC disabled
(ATC OFF):
For manual temperature compensation you can set the process and calibration
temperatures. This allows calibration at a temperature other than the process
temperature. Example: setting a calibration temperature of 25°C lets you calibrate
using standard solutions at 25°C, even if your process temperature is different from
25°C.
4.
Setting the process temperature: press the
▲
or
▼
key
to adjust the process
temperature (upper value, setting range: –10.0 to 125.0°C / 14.0 to 257.0 °F).
Press the
ENT key
to confirm process temperature.
5.
Setting the calibration temperature: press the
▲
or
▼
key
to adjust the calibration
temperature (upper value, setting range: – 10.0 to 125.0°C / 14.0 to 257.0 °F).
Press the
ENT key
to confirm the calibration temperature.
−
ATC enabled
(ATC ON):
6.
Setting the temperature offset: The display shows the current temperature offset (upper
value) and the temperature currently measured (lower value). Compare the current
temperature displayed on the transmitter to a thermometer known to be accurate. Note
the correct temperature value. Press the
▲
or
▼
key to adjust the lower value. The
upper number indicates the offset value. You can offset the temperature by up to
±10°C/±18°F. Press the ENT key to confirm your setting.
Continue with additional Setup procedures, or return to the Measurement mode by pressing
the
▲
and
▼
key (escape) simultaneously.
Summary of Contents for Aquacon 4000
Page 1: ...Instruction Manual Aquacon 4000 Conductivity Controller Transmitter 01 2004...
Page 11: ...Instruction Manual Aquacon 4000 6...
Page 15: ...Instruction Manual Aquacon 4000 10...
Page 27: ...Instruction Manual Aquacon 4000 22...
Page 45: ...Instruction Manual Aquacon 4000 40...
Page 51: ...Instruction Manual Aquacon 4000 46 Panel mounting version view from top...
Page 58: ......