FONA XPan 3D Plus
21
NOTE
The pre-programmed values of technique factors are factory programmed. Different
values can be loaded if needed using the available on board programming functionality.
Refer to §8, Programming for details.
Setting Tube Voltage
Key INCREASE at the left of the display to manually raise kV level
Key DECREASE at the left of the display to manually lower kV level
The tube voltage can be set from 61 to 85 kV in steps of 3 kV.
61
64
67
70
73
76
79
82
85
Setting Tube Current
Key INCREASE at the right of the display to manually raise mA level.
Key DECREASE at the right of the display to manually lower mA level.
The tube current can be set from 4 to 10 mA.
4.0
5.0
6.3
8.0
10
F. TEST Mode without Radiation (optional)
READY GREEN
LIGHT ON
Press RETURN Arm Movement to move the arm from PATIENT ENTRY position to
START position, ready to start the exposure.
When the unit reaches the START position the green light of the READY indicator on the
control panel is turned ON.
With unit in START position and the READY light ON, TEST mode can be activated to
run the unit without radiation.
On the display are shown:
x
The selected program number
x
The message “Test Mode” instead of values of kV and mA
x
A green light is also activated, above the key button
Starting the unit with the hand-switch allows for rotation of the arm according to the
program selected.
When the arm is returned to PATIENT ENTRY position test mode is terminated and the
units enters normal mode.