Calibration
6.2.2
Fully automatic calibration
If required, the stations equipped with the peristaltic pumps can
be all-automatically calibrated once a day (see
). If the calibration value determined via the auto-
matic calibration differs considerably from the value determined
so far, the calibration value until now will persist and the warning
autom. cal.
will be given. For each feeding station a tolerance
value can be defined separately. This value determines the max-
imum percentage the value determined via the automatic cali-
bration is allowed to differ from the calibration value until now in
order to be still considered valid, thus being taken over as the
new value.
1.
>
calibration
>
F-station 3
2. In
tolerance
enter the desired percentage of the tolerance
value.
6.2.3
Calibrating at the feeding station
1. Remove the suction hose from the teat at the feeding station
to be calibrated and hold the hose opening into the graduated
cylinder.
2. Press the calibration button at the IFS, which is allocated to
the feeding station. The feed is now dispensed via the suction
hose into the graduated cylinder. Keep the button pressed un-
<F-station 3>
start?
measured?
set qty: 500 ml
pulses: 110
tolerance: [15]%
date: 29.04.09
Default value:
15 %
Allowed range:
5 to 50 %
Note
: If the value determined via the automatic calibra-
tion is undone, the warning
autom. cal.
will be given
Summary of Contents for TAK5-CH1-25
Page 38: ...38 Operation ...
Page 54: ...54 Start up ...
Page 70: ...70 Setup ...
Page 84: ...84 Device data ...
Page 90: ...90 Calibration ...
Page 150: ...150 Cleaning ...
Page 170: ...170 Animal control ...
Page 192: ...192 Failures and warnings ...
Page 196: ...196 Care and maintenance plan Routine tasks ...
Page 215: ......
Page 218: ...218 Annex 16 3 5 Template for individual feeding plan ...
Page 219: ...Annex 219 16 4 Basic principle of interval feeding ...
Page 222: ...222 Annex ...