![Focustronic Alkatronic User Manual Download Page 20](http://html1.mh-extra.com/html/focustronic/alkatronic/alkatronic_user-manual_2314617020.webp)
20
measurement is going to be performed, and on the second line, if any actions were
taken based on the last measured value (Socket o
ff
, Dosed xx.xx ml, dKH OK).
•
Last measured value
•
When next measurement will be
•
If any actions were taken based
on the last measured value.
•
dKH OK
•
Dosed xx.xx ml
•
Socket o
ff
When making changes in settings, LCD screen will shortly show these settings.
You
will of course always see the settings in the App, as the App and
Alkatronic
remembers all settings, even after a power-o
ff
.
•
Low reference value
•
High reference value
•
Aquarium Volume
•
Measure frequency (distance
between 2 measurements in hour).
Measurement postponed function
The scheduled measurement is always starting at XX:00. That means that if you
happen to play with the app
, like sending commands to Alkatronic (like doing a
calibration or something else)
within 10 minutes before a scheduled
measurement should be performed
, then this scheduled measurement will not
start, but is blocked, both for convenience and for safety reason. In that case, the
machine will postpone the scheduled measurement one hour, to XX+1:00, and will
in LCD-display show, after your are done with the app-commands, :”
Measure
postponed to XX+1:00
”. You can still finish your commands to the app until XX:30.
After that time, XX:30, in ”postponed mode”,
the machine will not react to any
commands from app, to ensure the measurement, now XX+1:00, for sure will
be performed
.
Export Data
In the main menu, you have a button called
”
Export Data
”. When clicking on this, your
mail-client will be opened and include the
measurements from the history, in that mail.