![FMS EMGZ Series Operating Manual Download Page 38](http://html1.mh-extra.com/html/fms/emgz-series/emgz-series_operating-manual_2313008038.webp)
Operating Manual EMGZ611A...614A
38
14.3
Technical Data
Number of measuring
points
1...4 (depending on device type)
Connection of force
sensors
1 or 2 parallel wired force sensors of 350
for
each channel
Excitation of force sensors
5VDC (default), or 10VDC
(with automatic current control)
Input signal voltage
0...9mV (max. 12.5mV) or 0...18mV (max. 25mV)
(depending on force sensor excitation)
Resolution A/D converter
±8192 Digit (14 Bit)
Measuring error
<0.05% FS
Cycle time
2ms
Operation
4 keys, 4 LED’s, LCD display 2x16 characters (8mm
height)
Analog output channel
1...4
0...10V (default) / ±10V
and 0...20mA (default) / 4...20mA (12 Bit)
Digital output 1...4
(freely programmable)
Open collector, max. 10mA,
galvanically isolated, with recovery diode
Digital input 1...4
24VDC, galvanically isolated
(signal must be on for min. 100ms)
Interface RS232
Optional
Interface PROFIBUS
PROFIBUS DP (EN50170), optional
Interface CAN-Bus
Optional
Interface DeviceNet
Optional
Power supply
24VDC (18...36VDC) / 10W (max. 1A)
Temperature range
0...45°C (32...113°F)
Weight
1.5kg (3.35lbs)
Summary of Contents for EMGZ Series
Page 6: ...Operating Manual EMGZ611A 614A 6 Variant or option in italic text...
Page 11: ...Operating Manual EMGZ611A 614A 11 6 Dimensions fig 3 Dimensions E611002e...
Page 29: ...Operating Manual EMGZ611A 614A 29 10 Serial Interface RS232 Optional...
Page 35: ...Operating Manual EMGZ611A 614A 35 13 Interface CAN Bus Optional...