Service Information

10

Low Battery Indicator

To verify the correct operation of the low battery indicator:

1.

Remove the Tester batteries.

2.

Set the dc power supply to 3.0 V. Apply this voltage to the Tester battery terminals and observe the correct polarity.

3.

Set the Tester to the resistance measurement mode.

4.

Verify that four bars (full battery) show inside the battery indicator on the Tester display.

5.

Set the calibrator to 100 

Ω

. Apply this resistance to the Tester probes.

6.

Reduce the dc power supply voltage down to 2.3 V.

7.

Wait approximately 10 seconds for the battery indicator to update.

8.

Verify that 0 bars (empty battery) show inside the battery indicator on the Tester display.

9.

Verify that the Tester display reading is 97 

Ω

 to 103 

Ω

.

10.  Turn off the Tester.
11.  Disconnect the dc power supply leads from the Tester.
12.  Disconnect the calibrator from the Tester.
13.  Reinstall the Tester batteries. 

Battery Door Circuit 

To verify the battery door circuit operates correctly:

Note

During this test, do not use a sharp test probe to contact the battery door pad. The pad can be 
easily damaged. 

1.

Set the 8846A to ACI and 1 mA range.

2.

Set the 8846A resolution to LOW (MEAS SETUP > RESOLUTION > LOW). This hides the two least significant digits that are 
noisy.

3.

Set the calibrator to 20 V, 55 Hz.

4.

Connect the test equipment as follows:

a. 8846A 400 mA jack to the battery door probe dock spring.

b. calibrator NORMAL LO to the 8846A LO jack (current).

c. calibrator NORMAL HI to the battery door pad.

5.

Set the calibrator to OPERATE.

6.

Verify that the 8846A reads between 21 

μ

A and 15 

μ

A (0.0210 to 0.0150).

7.

Change the calibrator frequency to 1 kHz and confirm the calibrator is in OPERATE mode.

8.

Verify that the 8846A reads between 112 

μ

A and 68 

μ

A (0.1120 to 0.0680).

9.

Set the Calibrator to STANDBY.

10. Disconnect the battery door from the test instruments and reinstall it onto the Tester.

Summary of Contents for T6-1000

Page 1: ...8 2018 Fluke Corporation All rights reserved All product names are trademarks of their respective companies Specifications are subject to change without notice T6 1000 T6 600 Electrical Tester Service...

Page 2: ...ITY OR FITNESS FOR A PARTICULAR PURPOSE FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE OR THEORY Since s...

Page 3: ...ts 5 Annunciator and Hazardous Voltage Indicator 5 Voltage Function 6 FieldSense Current Function 7 FieldSense Voltage Function 9 Resistance and Continuity Functions 9 Low Battery Indicator 10 Battery...

Page 4: ...ii T6 1000 T6 600 Service Information...

Page 5: ...world 1 425 446 5500 Or visit Fluke s website at www fluke com To register your product visit http register fluke com To view print or download the latest manual supplement visit http us fluke com us...

Page 6: ...rated voltage between the terminals or between each terminal and earth ground Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product probe or accessory...

Page 7: ...s applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Category IV is applicable to test and measuring circuits con...

Page 8: ...s Caution This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments Korea KCC Class A Equipment Industrial Broa...

Page 9: ...in Table 3 Table 3 Annunciator Tests 4 Set the calibrator to the voltage and frequency listed in Table 4 Step 1 5 Apply the voltage to the Product and verify the readings are within the display limits...

Page 10: ...readings for the remaining steps in Table 6 T6 600 T6 1000 Step Calibrator Output Display Limits AC Annunciator ON Calibrator Output Display Limits AC Annunciator ON 1 200 V 60 Hz 195 V to 205 V 200...

Page 11: ...the SHIFT yellow button on the Product to switch to AC Current mode and Frequency mode T6 1000 only 2 Connect the 10 50 turn coil to the calibrator AUX HI and AUX LO jacks 3 Connect the BLACK lead of...

Page 12: ...hat the Tester reading is within the display limits 11 Apply the test values shown in the remaining steps of Table 8 and verify the Tester readings are within the display limits Table 8 FieldSense Hig...

Page 13: ...the tester to switch to the AC voltage mode 8 Place the Product current fork into position on the PRV240FS 9 Place the BLACK test lead of the tester into the Test Lead Ground Contact of the PRV240FS...

Page 14: ...sconnect the calibrator from the Tester 13 Reinstall the Tester batteries Battery Door Circuit To verify the battery door circuit operates correctly Note During this test do not use a sharp test probe...

Page 15: ...ts Mild soap may be used Battery Replacement W Caution For safe operation and maintenance of the Product Remove batteries to prevent battery leakage and damage to the Product if it is not used for an...

Page 16: ...able 10 To order parts and accessories see How to Contact Fluke Table 10 Replacement Parts Ref Description Fluke Part or Model Number Battery Door 4944370 Battery Door Screw 1618579 Test Lead Assembly...

Reviews: