Using the MS190 Function
2
2-3
The screen shows :
•
Top left reading, A (mA,
µ
A, etc.):
the difference between the current amplitude measured
on the left cursor and measured on the right cursor:
•
Top right reading, mAs (As,
µ
As, etc.):
the integrated current over the time between the
cursors.
12
Highlight the left cursor.
13
Move the left cursor to the
desired position on the waveform.
14
Highlight the right cursor.
15
Move the right cursor to the
desired position on the waveform.
Extended Offset
The MS190 extended offset function enlarges the vertical
trace positioning range for input A and input B to 16
divisions (maximum). This allows you to zoom-in vertically
to study small details of the signal.
To choose the extended offset range for input A do the
following:
1
Display the
INPUT A
key labels.
2
Open the
INPUT A
menu
3
(2x)
Jump to Offset range.
4
Select Extended and accept the
extended offset range